background image

  HVRGRN8BLWS*1C     HVRGRN8BLWSNF*1C

1000001740  (Rev. P - 03/23) 

Page 7

VERIFY CONTROL BOARD  SOFTWARE

1) To verify the software program of the control board the unit will  

    need to be shut down and restarted. The chiller (if present) does 

    not need to be shut down and restarted.

2) The units lower panel must be open to access the power cord and 

    wall outlet.

3) Shut down the unit by unplugging the power cord from the wall 

    outlet. 

4)  Restart the unit by plugging the power cord back into the wall 

    outlet.

5) Upon start up, the bottle count display will show the software 

    designation of BF11 or BF12.

ACCESSING THE PROGRAMMING BUTTON

1)   To access the program button, remove the top cover of the bottle- 

  filler. Remove the two (2) screws holding top cover to bottle-filler 

  with a 5/32” allen wrench. Remove top cover. Do not discard

  mounting screws, they will be needed to reinstall the top cove after  

  programming operations are completed. The programming button  

  is located at the top right side of the unit on the control board.  

 

NOTE: When applicable, there is also an alternate reset button  

  located on the lower part of the water cooler. After removing

  the bottom cover, the reset button will be located on the left 

  side of the cooler, mounted on the side panel support.

RESET THE FILTER MONITOR

1) Instructions apply to filtered units only.

2) Depress the program button for approximately 2 seconds until 

    the display changes then release. The display will change and 

    scroll through two messages:

       “RST FLTR” – Reset Filter Monitor

     “SETTINGS” – System Settings Sub Menu

    If the program button is not pushed again the display will scroll 

    through the two messages above for three cycles and then default 

    back to bottle count and be back in run mode.

3) When the display changes to “RST FLTR”, depress the button 

    again. The display will change to show “FLTR =”. Depress the

    button again and the display will show “FLTR =0”

4) The Green LED should be illuminated indicating that the visual 

    filter monitor has been reset.

 SETTING RANGE OF THE IR SENSOR WHERE APPLICABLE

1) Depress the program button for approximately 2 seconds until 

    the display changes then release. The display will change and 

    scroll through two messages:

       “RST FLTR” – Reset Filter Status LED

     “SETTINGS” – System Settings Sub Menu

    If the program button is not pushed again the display will scroll 

    through the two messages above for three cycles and then default 

    back to bottle count and be back in run mode.

2) When the display changes to “SETTINGS”, depress the button 

    again. The display will change to show

      “RNG SET” -  Range set for IR sensor.

      “UNIT TYP” - Type of unit (REFRIG or NON-RFRG)

      “FLT SIZE” - Select filter capacity

      “RST BCNT” - Reset bottle count

3)  When display shows “RNG SET” push program button once the 

    display will show current value (can be 1 – 10) e.g. “RNG = 3”.

4) Once display shows current value push the program button to 

    scroll through value of 1 – 10. Select the desired range setting, 

  "1" being closest to sensor and "10" being farthest away.

5) Once range is selected allow approximately 4 seconds to pass and

    then the display will go back to bottle counter and be in run mode.

6)  Test bottle filler by placing bottle or hand in front of sensor to 

    make sure water is dispensed.

SETTING UNIT TYPE

1) Depress the program button for approximately 2 seconds until the

    display changes then release. The display will change and scroll 

    through two messages:

       “RST FLTR” – Reset Filter Status LED

      “SETTINGS” – System Settings Sub Menu

    If the program button is not pushed again the display will scroll 

    through the two messages above for three cycles and then default 

    back to bottle count and be back in run mode.

Continued from below:

2)  When the display changes to “SETTINGS”, depress the button again.  

      The display will change to show

       “RNG SET” -  Range set for IR sensor.

       “UNIT TYP” - Type of unit (REFRIG or NON-RFRG)

       “FLT SIZE” - Select filter capacity

       “RST BCNT” - Reset bottle count

3)  When display shows “UNIT TYPE” push program button once the

    display will show current value.  Can be REFRIG or NON-RFRG

4) Push button once to change value. Once value is    selected the display 

    will show the new value.  (Can be REFRIG or NON-RFRG)  

     

“REFRIG“ - stands for refrigerated product. In this setting the flow rate is 

    estimated at 1.0 gallon per minute.

     “NON-RFRG“ - stands for nonrefrigerated product. In this setting the  

    flow rate is estimated at 1.5 gallons per minute.  Both “REFRIG“ and 

    “NON-RFRG“ simulate   1 bottle equal to 20 oz.

5)  

Allow approximately 4 seconds to pass and the display will return to    

    bottle counter and be in run mode.

    

 RESETTING BOTTLE COUNT

1) Depress the program button for approximately 2 seconds until the

    display changes then release. The display will change and scroll

    through two messages:

      “RST FLTR” – Reset Filter Status LED

      “SETTINGS” – System Settings Sub Menu

    If the program button is not pushed again the display will scroll through  

    the two messages above for three cycles and then default back to bottle 

    count and be back in run mode.

2)  When the display changes to “SETTINGS”, depress the button again. 

     The display will change to show:

      “RNG SET”-  Range set for IR sensor.

      “UNIT TYP” - Type of unit (REFRIG or NON-RFRG)

       

“FLT SIZE” - Select filter capacity

      “RST BCNT” - Reset bottle count

    If the button is not pushed again the display will scroll through the four 

    messages above for three cycles and return to run mode.

3)  When display shows “RST BCNT” push program button once the

    display will show current value, e.g. “0033183”.

4) Once display shows current value push the program button once more 

    to reset back to 0. The display will show BTLCT = 0 for approximately 2 

    seconds and then return to run mode showing 00000000 bottles. 

 

NOTE: Once the bottle count is reset to zero there is no way to 

  return to the previous bottle count.

5)  Testing the bottle counter:

     REFRIG units: Place bottle or hand in front of sensor for approximately 

  9 seconds to see bottle counter count 00000001, 

     (This is based on filling a 20 oz. bottle).

     NON-RFRG units: Place bottle or hand in front of sensor for 

  approximately 6 seconds to see bottle counter count 00000001,

     (This is based on filling a 20 oz bottle).

    

 SETTING FILTER CAPACITY

1) Depress the program button for approximately 2 seconds until the

    display changes then release. The display will change and scroll through 

    two messages:

      “RST FLTR” – Reset Filter Status LED

      “SETTINGS” – System Settings Sub Menu

    If the program button is not pushed again the display will scroll through 

    the two messages above for three cycles and then default back to bottle 

    count and be back in run mode.

2)  When the display changes to “SETTINGS”, depress the button again. 

     The display will change to show:

      “RNG SET“-  Range set for IR sensor.

      “UNIT TYP“ - Type of unit (REFRIG or NON-RFRG)

      

 “FLT SIZE” - Select filter capacity

      “RST BCNT“ - Reset bottle count

    If the button is not pushed again the display will scroll through the four 

    messages above for three cycles and return to run mode.

3)  When display shows “FLT SIZE” push program button once. The display 

     will show current value. Can be 3000GAL or 6000GAL.

4)  Push program button again to display the desired “FLT SIZE”. 

5)  Allow approximately 4 seconds to pass and the display will return to 

     bottle counter and be in run mode.

BF11 - BF12 PROGRAM

SETTING THE CONTROL BOARD

Summary of Contents for Hydroboost HVRGRN8BLWS 1C Series

Page 1: ...Cooler Owners Manual Estaci n de llenado de botella Hydroboost y enfriador Station de remplissage de bouteille Hydroboost et refroidisseur WARNING Cancer and Reproductive Harm www P65Warnings ca gov...

Page 2: ...5 19 12 Pictured is unit only without bottle filler Uses HFC 134A refrigerant Usa refrigerante HFC 134A Utilise du fluide frigorig ne HFC 134A Note Danger Electric shock hazard Disconnect power befor...

Page 3: ...A DE PARED L O D de 3 8 d EMPLACEMENT DE PROVISION D EAU RECOMMANDE LE TUBE DE CUIVRE DE UNPLATED CONNECTE STUB AVEC ETEINT PAR LES AUTRES 3 dans 76 mm le MAXIMUM HORS DU MUR B RECOMMENDED LOCATION FO...

Page 4: ...e posterior del enfriador como se indica en la figura 4 5 Afloje los dos 2 tornillos que sostienen la parte inferior del panel en la parte inferior de la base del enfriador y los dos 2 tornillos en la...

Page 5: ...S 1C HVRGRN8BLWSNF 1C 1000001740 Rev P 03 23 Page 5 Fig 5 27 2 2 2 26 2 2 1 PUSH BUTTON MECHANISM MECANISMO DE BARRA DE EMPUJE M CANISME DU BOUTON POUSSOIR 2 28 9 17 Fig 6 Fig 7 16 4B 24 20 24 25 4A 2...

Page 6: ...om the wall mounting plate installed on wall Make sure round boss in gasket fits in hole of basin See Fig 12 9 Once Bottle Filler is installed on wall plate tabs water line and power cord are installe...

Page 7: ...ay will scroll through the two messages above for three cycles and then default back to bottle count and be back in run mode Continued from below 2 When the display changes to SETTINGS depress the but...

Page 8: ...3000 Gallon Kit filtre raccord de t te comprend Raccords John Guest Assembl e filtre t te Support inclut John Guest vis DESCRIPTION FILTER PARTS LIST See Fig 15 LISTA DE PIEZAS DEL FIL TRO Vea la Fig...

Page 9: ...83C 5 16 1 4 Pack de 3 Kit Tee 1 4 x 1 4 x 3 8 Pack de 3 Kit 10 Torx vis T 25 bits Kit de Filtrage d Eau Si Fourni DESCRIPTION Panel Bottom Dispenser Kit Push Button Assy Button Sleeve Nuts Spacer Han...

Reviews: