1: Setup
•
One file has the calibration numbers when calibrating the spectral response of your system
with the lamp and a
CC-3 cosine-corrector and fiber
. The file name contains the lamp serial
number, followed by
CC3.LMP.
2. Copy these two files into your OOIIrrad Irradiance Software directory.
3. Start OOIIrrad Irradiance Software and place the software in
Scope Mode.
4. Choose
the
Select Lamp
option from the Lamp menu. A window opens and prompts you to
choose the Lamp Calibration Report file that reflects your optical setup (see Step 1 for details).
5. Choose
Configure Fibers
from the Spectrometer menu. Enter the fiber diameter values for each
channel in your setup. If using a bare fiber, enter the fiber's diameter here. If using a fiber with a
CC-3 cosine corrector, enter 3,900.
Note:
Whatever optical setup you wish to use for your application, you must also use for
calibrating the spectral response of your system. For example, if you are going to use a
200 µm fiber with a CC-3 cosine corrector for your application, you must use the same
200 µm fiber and CC-3 for calibrating the spectral response of your system.
6. Enter the data acquisition parameters in the appropriate boxes located to the right of the displayed
spectrum. The values for Scans to Average and Smoothing Size must be the same for your
reference, dark and irradiance scans. You may use a different integration period for your
reference scan, but you must use the same integration period for your dark and irradiance scans.
4
000-10000-060-02-0505