Halma Ocean Optics HL-2000-CAL Installation And Operation Manual Download Page 12

1: Setup 

 

• 

One file has the calibration numbers when calibrating the spectral response of your system 
with the lamp and a 

CC-3 cosine-corrector and fiber

. The file name contains the lamp serial 

number, followed by

 CC3.LMP.

 

2.  Copy these two files into your OOIIrrad Irradiance Software directory. 

3.  Start OOIIrrad Irradiance Software and place the software in 

Scope Mode.

 

4. Choose 

the 

Select Lamp 

option from the Lamp menu. A window opens and prompts you to 

choose the Lamp Calibration Report file that reflects your optical setup (see Step 1 for details). 

5. Choose 

Configure Fibers

 from the Spectrometer menu. Enter the fiber diameter values for each 

channel in your setup. If using a bare fiber, enter the fiber's diameter here. If using a fiber with a 
CC-3 cosine corrector, enter 3,900.  

Note:  

Whatever optical setup you wish to use for your application, you must also use for 
calibrating the spectral response of your system. For example, if you are going to use a 
200 µm fiber with a CC-3 cosine corrector for your application, you must use the same 
200 µm fiber and CC-3 for calibrating the spectral response of your system. 

6.  Enter the data acquisition parameters in the appropriate boxes located to the right of the displayed 

spectrum. The values for Scans to Average and Smoothing Size must be the same for your 
reference, dark and irradiance scans. You may use a different integration period for your 
reference scan, but you must use the same integration period for your dark and irradiance scans. 

4

 

000-10000-060-02-0505 

Summary of Contents for Ocean Optics HL-2000-CAL

Page 1: ...34698 Phone 727 733 2447 Fax 727 733 3962 8 a m 8 p m Mon Thu 8 a m 6 p m Fri EST Ocean Optics B V Europe Geograaf 24 6921 EW DUIVEN The Netherlands Phone 31 0 26 3190500 Fax 31 0 26 3190505 E mail In...

Page 2: ...2005 Ocean Optics Inc All rights reserved No part of this publication may be reproduced stored in a retrieval system or transmitted by any means electronic mechanical photocopying recording or otherwi...

Page 3: ...anty and create an unsafe operating environment 2 Dangerous voltages are present in this device There are NO user serviceable parts inside 3 Only allow qualified personnel to service this unit 4 Do no...

Page 4: ...Important Safety Notices B 000 10000 060 02 0505...

Page 5: ...ated Documentation iii Upgrades iv Chapter 1 Setup 1 Overview 1 Unpacking the HL 2000 CAL 2 Contents 2 Connecting a Spectrometer to the HL 2000 CAL 3 Configuring the OOIIrrad Software 3 Chapter 2 HL 2...

Page 6: ...Table of Contents ii 000 10000 060 02 0505...

Page 7: ...ing environment specifications as well as other physical details of the product Chapter 3 Calibration Provides instructions for calibrating the spectrometer connected to the HL 2000 CAL Light Source P...

Page 8: ...change or an upgrade to your system To facilitate these changes you must first contact Customer Support and obtain a Return Merchandise Authorization RMA number Please contact an Ocean Optics Applicat...

Page 9: ...e absolute spectral response of your system Note If you have a spectrometer setup that is highly sensitive you may not be able to use the HL 2000 CAL as a calibration source Before using the HL 2000 C...

Page 10: ...a clean laboratory environment see Operating Environment Contents Your DH 2000 CAL package should contain the following DL 2000 CAL unit 1 Switching AC adapter for stabilizing power 1 Power cord for s...

Page 11: ...f the HL 2000 CAL with an Allen wrench Screw the other end of the fiber into the SMA connector of the spectrometer 4 Plug the switching AC adapter the black rectangle shaped box into the back of the H...

Page 12: ...our setup If using a bare fiber enter the fiber s diameter here If using a fiber with a CC 3 cosine corrector enter 3 900 Note Whatever optical setup you wish to use for your application you must also...

Page 13: ...tion Heat Should be situated away from any device that emits excessive heat Object and Liquid Entry Should be positioned so that objects do not fall on top of the unit Additionally ensure that no liqu...

Page 14: ...2 HL 2000 CAL Specifications 6 000 10000 060 02 0505...

Page 15: ...ctrometer for a DARK scan 4 Completely block the light path to your spectrometer by inserting an opaque object into the lamp s filter slot if possible do not turn off the light source Taking a dark sp...

Page 16: ...3 Calibration 8 000 10000 060 02 0505...

Page 17: ...nce iii purpose iii summary iii O OOIIrrad software configuring 3 operating environment 5 P package contents 2 product related documentation iii S setup 1 specifications 5 specifications table 5 spect...

Page 18: ...Index 10 000 10000 060 02 0505...

Reviews: