22
Troubleshooting and Maintenance
Issue:
My pressure controller disagrees with another device I have in line.
Action:
Pressure devices can normally be compared against one another provided there are no leaks between the two devices.
Another possibility is an improper tare error (
).
Issue:
Can I use the controller with other gases or liquids?
Action:
Yes for gases, maybe for liquids. Your pressure controller is designed to operate independent of the media being
used. One thing to check before changing gases or liquids is the chemical and material compatibility of the gas being
used with the wetted materials inside the controller. We also recommend contacting support (
) before
switching over to liquid pressure control.
Serial Communications
Issue:
I can’t communicate to the device when it is connected to my PC.
Action: 1.
Make sure the baud rate your software and COM port required is the one your pressure controller is using (
MENU
→
SETUP
→
RS-232 Serial
or
RS-485 Serial
→
Baud Rate
).
2.
Check the unit ID (
MENU
→
SETUP
→
RS-232 Serial
or
RS-485 Serial
→
Unit ID
) to make sure you are addressing
it properly with your serial commands.
3.
Check the pinout (common pinouts are listed starting on
4.
Make sure the COM number matches the one your software is using to connect to the pressure controller.
5.
On the external serial communications device (computer, PLC,
etc.
), be sure that the pressure control (handshaking)
settings are set as on
.
Still experiencing issues? Please contact support. See “Contact Information” on (
).
Maintenance
Cleaning
This device requires minimal maintenance. If necessary, the outside of the device can be cleaned with a soft dry cloth.
Avoid excess moisture or solvents.
The primary causes of damage and/or long-term inaccuracy in these devices are contamination and/or corrosion damage.
Fluid should be filtered for particulates or biological materials that may grow in the device.
Recalibration
The recommended period for recalibration is once every year. A label located on the back of the device lists the most
recent calibration date. This date is also stored inside your device and is visible by selecting
MENU
→
ABOUT
→
About Device
.
When it is time for your device’s annual recalibration, contact support (
) with your device’s serial number and
your contact information.