Page 12 of 24
V2.0
every 4 minutes, and the sample time indicator will display 4. Note, a manual
measurement can
always
be made by pressing the
ENTER
button.
Sound Indicator:
The second icon in the Indicator Display is the sound ON/OFF icon.
Note, this options refers to the sounds made when interface or keypad buttons are
pressed. It does not affect the alarm or the startup sounds. In
has its sound turned on. If the sounds were turned off, an X would appear next to the
sound icon.
Alarm Indicator:
The third icon in the Indicator Display is the alarm ON/OFF icon. In
the alarm icon shows that the alarm is turned on. To set the alarm, go to the
Settings screen and modify the alarm level. To turn the alarm off, set its level to 0 PPM.
If the alarm were turned off, an X would appear next to the alarm icon.
Battery Status
:
The battery indicator displays the battery
’s remaining charge
graphically. Three bars represent 100% of charge in the battery; two bars 75%; one bar
50%. No bars signifies a low battery status and simultaneously the alarm will buzz as a
warning. Charging of the battery is necessary at this level and after a few seconds of
audible warnings the instrument will shut off.
Measurement Gas:
It shows a gas type that Formaldehyde is being sampled.
Running Indicator:
When the unit is measuring, the running icon will be displayed and
the pump will be operating. When the unit is idle (not measuring), the running icon will
not be displayed and the pump will be off.
Temperature:
The current ambient temperature is displayed in units of degrees Celsius.
This measurement is not standard and requires the optional temperature sensor probe
to function. If no probe is installed, the unit will display: - - ºC.