日本文
CAUTION
• Do not bring the iron tip into contact with the
resin components including terminals and
slide pin of the tester.
• The measuring point of the sensor generally
undergoes degradation as a result of
repeated measurement activities. It is
recommend that the sensor be replaced
every 50 measurements as a guideline to
ensure measurement accuracy.
• If the terminals are contaminated with the
soldering flux, wipe them clean with alcohol.
Do not use thinner or benzin for cleaning.
• Please read when the temperature
stabilizes.
n
MAX HOLD function
When pressing the
MAX HOLD
button, "MAX
HOLD" is displayed at the lower right of
the LCD. As long as
MAX HOLD
appears, the
maximum temperature will stay displayed.
OPERATION
The
MAX HOLD
button provides two additional
functions: The max. temperature update
function when quickly pressing the button
and the MAX HOLD cancellation function
when pressing the button longer.
MAX HOLD
MAX HOLD
MAX HOLD
MAX HOLD
Press
less than 1 second
MAX HOLD
Press the button for
1 second or more
MAX HOLD
Press the button for
less than 1 second
Max. temperature is updated.
"MAX HOLD"
disappears
indicating the
function is
cancelled.
• Quickly pressing the
MAX HOLD
button for less
than one second, with "MAX HOLD"
displayed, updates the maximum
temperature. See the figure to the right.
• With "MAX HOLD" displayed, pressing the
button for longer than one second cancels the
MAX HOLD function. See the figure to the right.
— NOTE —
• Turning off the power always cancels the
MAX HOLD function.
• The MAX HOLD function can only be used
during the temperature measurement.
13
2. Insert the power cord into the outlet at the
back of the body and turn the power switch on.
• Be sure to insert the power plug into a
grounded (3-hole) outlet.
• Power will be supplied through the
receptacle on the HAKKO FG-101 only
when the power switch is turned on.
(1) Measuring the tip temperature
1. Press the SELECT button to light up the
mode indicator lamp of "TEMP."
2. Place the tip coated with fresh solder on the
measuring point (Refer to the figure at right).
Tip
Fresh solder
Measuring point
CAUTION
Do not directly contact the HAKKO FG-101
with hot air (HAKKO FR-810, etc.), during
measurement. Direct contact with hot air can
damage the HAKKO FG-101.
CAUTION
Handle the sensor with care. Tough handling
may break the CA sensor wire as it is as thin as
0.2 mm in diameter.
— NOTE —
Use the iron tip coated with fresh solder when
performing mesurement to ensure contact
between the temperature sensor or the
conduction plate and the tip.