 CAUTION

• Do not bring the iron tip into contact with the 

resin components including terminals and 

slide pin of the tester.

• The measuring point of the sensor generally 

undergoes degradation as a result of 

repeated measurement activities. It is 

recommend that the sensor be replaced 

every 50 measurements as a guideline to 

ensure measurement accuracy. 

• If the terminals are contaminated with the 

soldering flux, wipe them clean with alcohol. 

Do not use thinner or benzin for cleaning. 

• Please read when the temperature 

stabilizes.

MAX HOLD function

When pressing the 

MAX HOLD

 button, "MAX 

HOLD" is displayed at the lower right of 

the LCD. As long as  

MAX HOLD

  appears, the 

maximum temperature will stay displayed. 
OPERATION

The 

MAX HOLD

 button provides two additional 

functions: The max. temperature update 

function when quickly pressing the button 

and the MAX HOLD cancellation function 

when pressing the button longer.

MAX HOLD

MAX HOLD

MAX HOLD

MAX HOLD

Press
less than 1 second

MAX HOLD

Press the            button for
1 second or more

MAX HOLD

Press the            button for
less than 1 second

Max. temperature is updated.

"MAX HOLD"

disappears 

indicating the 

function is 

cancelled.

• Quickly pressing the 

MAX HOLD

  button for less 

than one  second, with "MAX HOLD" 

 displayed, updates the  maximum 

temperature.  See the figure to the right.

• With "MAX HOLD" displayed,  pressing the           

button for  longer than one second cancels  the 

MAX HOLD function.  See the figure to the right.

— NOTE —

• Turning off the power always cancels the 

MAX HOLD function.

• The MAX HOLD function can only be used 

during the temperature measurement. 

13

2. Insert the power cord into the outlet at the 

back of the body and turn the power switch on.

• Be sure to insert the power plug into a 

grounded (3-hole) outlet.

• Power will be supplied through the 

receptacle  on the HAKKO FG-101 only 

when the power switch is turned on. 

(1) Measuring the tip temperature

1. Press the SELECT button to light up the 

mode indicator lamp of "TEMP."

2. Place the tip coated with fresh solder on the 

measuring point (Refer to the figure at right).

Tip

Fresh solder

Measuring point

 CAUTION

Do not directly contact the HAKKO FG-101 

with hot air (HAKKO FR-810, etc.), during 

measurement. Direct contact with hot air can 

damage the HAKKO FG-101.

 CAUTION

Handle the sensor with care. Tough handling 

may break the CA sensor wire as it is as thin as 

0.2 mm in diameter.

— NOTE —

Use the iron tip coated with fresh solder when 

performing mesurement to ensure contact 

between the temperature sensor or the 

conduction plate and the tip.

Summary of Contents for FG101-16

Page 1: ...English 1 FG 101 1 1 1 10 1 1 1 1 AUTO ZERO MAX HOLD SELECT GND FG 101 SOLDERING TESTER FG 101...

Page 2: ...00 600 5 K CA 0 1 mV 0 40 mV AC 5 of reading 1 digit 0 1 0 40 5 of reading 1 digit LCD 3 1 2 2 MAX HOLD MAX HOLD 2 6 W 200 W 50 H 120 D mm 1 kg 0 40 80 RH IEC UL61010 1 1 191 212 212C 500 932 F 500 93...

Page 3: ...AX HOLD MAX HOLD MAX HOLD MAX HOLD MAX HOLD 1 1 MAX HOLD 1 MAX HOLD MAX MAX HOLD MAX HOLD MAX HOLD MAX HOLD MAX HOLD MAX 1 MAX HOLD MAX MAX 1 MAX HOLD 1 1 SELECT TEMP 2 3 2 FG 101 FR 810 FG 101 FG 101...

Page 4: ...English 3 SELECT OHM AUTO ZERO AUTOZERO mV OHM AUTO ZERO 0 0 0 0 0 0 AUTOZERO 4 2 1 FG 101 2 3 SELECT mV 4 AUTO ZERO AUTO ZERO 5 6 7 2...

Page 5: ...English 5 191 212C 10 A1310 3 C1541 A B A1557 B 3 FG 101 FG 101 B1752 B1754 B1950 B2468 125 V 5 A B2387 191 212 10 50 5...

Page 6: ...English 1 HAKKO FG 101 1 1 1 qty 10 1 1 1 1 1 1 AUTO ZERO MAX HOLD GND 6 HAKKO FG 101 SOLDERING TESTER FG 101 HAKKO FG 101 3...

Page 7: ...CA 0 1 mV 0 40 mV AC 5 1 0 1 0 40 5 1 LCD 3 1 2 2 MAX HOLD MAX HOLD 100 V 2 6 W 110 V 2 9 W 120 V 2 6 W 220 V 2 7 W 230 V 2 8 W 240 V 3 0 W 200 W 50 H 120 D mm 1 kg 0 40 max 80 RH 2 IEC UL 61010 1 1...

Page 8: ...LD MAX HOLD MAX HOLD MAX HOLD MAX HOLD MAX MAX HOLD MAX HOLD MAX HOLD MAX HOLD MAX HOLD 1 1 MAX 1 1 MAXHOLD MAX MAX HOLD MAX MAX 1 MAX HOLD MAX HOLD 8 2 3 HAKKO FG 101 1 1 TEMP 2 HAKKO FR 810 HAKKO FG...

Page 9: ...English 2 1 HAKKO FG 101 2 3 mV 4 AUTO ZERO AUTO ZERO 5 6 7 MIL STD 2000 1 2 3 9 3 OHM AUTO ZERO 3 AUTO ZERO mV OHM AUTO ZERO 0 0 0 0 0 0 AUTO ZERO HAKKO FG 101 GND GND...

Page 10: ...A1310 3 C1541 A B A1557 B 3 HAKKO FG 101 HAKKO FG 101 B3213 B3214 B1752 B1754 B1950 B1258 250 V 3 15 A B2468 125 V 5 A 191 212 10 B2419 B2421 B2422 B2424 220 V KTL 230 V CE B2425 230 V CE B2426 B2436...

Page 11: ...Ground terminal GND Protection plate Mode indicator lamp 11 NOTE The Multi adapter and the European adapter may not be included depending on the type of electrical connection used in your country Mul...

Page 12: ...12 F 5 C 10 F other than above Sensor K CA type thermocouple Voltage Resolution 0 1 mV Measurement Range 0 to 40 mV AC Precision 5 of reading 1 digit Resistance Resolution 0 1 Measurement Range 0 to 4...

Page 13: ...celled Quickly pressing the MAX HOLD button for less than one second with MAX HOLD displayed updates the maximum temperature See the figure to the right With MAX HOLD displayed pressing the button for...

Page 14: ...e soldering iron is a variable tip temperature model set temperature to the maximum 3 Press the SELECT button to light up the mode indicator lamp of mV 4 Press the AUTO ZERO button Refer to AUTO ZERO...

Page 15: ...ng pot C1541 Temperature Probe for hot air with Sensor A B A1557 Sensor B 3 Remove the standard sensor and connect the red connector of this option to the red terminal of the HAKKO FG 101 and the blue...

Page 16: ...33 http www hakko com sg E mail sales hakko com sg Please access to the following address for the other Sales affiliates http www hakko com HEAD OFFICE 4 5 Shiokusa 2 chome Naniwa ku Osaka 556 0024 JA...

Reviews: