Haier HXC-1308 Operation Manual Download Page 21

     

Maintenance and change of emergency battery of the recorder 

When the green LED of recorder flashes, there are two checking methods: 
1. Discharge the emergency battery. If the LED lamp is off and the recorder stops operation, it 
indicates that there is problem on the recorder main power (220V/50HZ). Please check whether 
there is power outage, short circuit or loose wiring. Install the emergency power when the recorder 
operates normally after maintenance of main power. If LED lamp stops flashing, it indicates that the 
emergency power is normal. If the LED lamp continues flashing, it indicates that power of 
emergency battery is low. Emergency battery needs to be changed until the LED green lamp stops 
flashing. 
2. Discharge the emergency power. LED lamp of power supply continues flashing and recorder 
operates normally, which indicates that the main power is normal and the power of emergency 
battery is so low that it needs to be changed until LED green lamp works and stops flashing. 
     Change Recording Paper 
Change of recording paper shall be conducted in the following steps: 
1. Find the compress button (3#) on the left corner and in front of recorder panel; 
2. Press the button (3#) for about 1 second until the recording pen starts moving to the left side; 
3. Loosen the middle bolt when the recording pen moves outside the recording paper. Replace the 
old paper with a new one. Align the time line on paper with the time groove on the panel (there is a 
small groove on left side of the panel); 
4. Tighten the central bolt again. Press the button-3# for about 1 second, to reset the recording pen
and start recording temperature; 
5. Examine that whether the recording pen operates well on the paper. If not, beam of the recording 
pen can be adjusted to make the nib contact with recording paper. (Note: Do not damage the nib 
and beam. If the adjustment is difficult, screw-driver can be used to discharge the beam. Install the 
beam after curving the beam through slight force. Repeat the above operations); 
6. Accurate recording of the recording pen shall be guaranteed. Calibrate accuracy of the recording 
pen after exchange of paper. The method is as follow: Press button (3#) until recording pen leaves 
the paper. And press button (3#) again until the pen returns to the paper. At this moment, the pen 
will pause at the most external temperature scaling line of the recording paper (this scaling line does 
not have to be marked with temperature value, but must be the one at the most external circle). If 
the pen does not pause at the above position, you can use 1# or 2# arrow key to adjust the position 
within 5 seconds, to align the nib to the scaling line at the most external circle. If position of the pen 
is not adjusted well within 5 seconds, please repeat Step 6. 

19

26

Alarm 

or 

safety 

Phenomenon 

Alarm indication 

Buzzer alarm 

High 

temperature 

Low 

temperature

Power cut 

The door is 

open slightly 

Sensor 

abnormal 

Alarm light flashes 

Alarm light flashes

The temperature display area 

displays the temperature in 

refrigerator for 60s, then stops 

displaying for 60s, and they will 

conduct alternatively 

After delay for 10min, alarm 

light flashes 

Pulse sound alarm 

Within 48h of 

power outage, 

pulse sound alarm 

occurs 

After delay for 

10min, pulse sound 

alarm occurs 

If the upper or lower 

temperature sensor 

perceives that the 

refrigerator temperature 

is higher than 6

 

Refrigerator is in power 

outage status 

The outside door is 

open slightly or fully 

Open circuit or short 

circuit occurs on upper 

temperature sensor 

If the upper or lower 

temperature sensor 

perceives that the 

refrigerator temperature 

is lower than 2

 

Open circuit or short 

circuit occurs on lower 

temperature sensor 

Open circuit or short 

circuit occurs on 

control sensor 

Open circuit or short 

circuit occurs on 

defrosting sensor 

The alarm light flashes and E1 

is displayed in temperature 

display area 

The alarm light flashes and E2 

is displayed in temperature 

display area 

The alarm light flashes and E3 

is displayed in temperature 

display area

The alarm light flashes and E4 

is displayed in temperature 

display area

The rechargeable 

battery voltage is too 

low or not connected. 

The alarm light is on constantly.

 “E5” is displayed in 

temperature display area for 3 

times and 3s each time. 

Safe operation 

Trigger remote 

alarm contact 

Uninfluenced 

Pulse sound alarm 

Pulse sound alarm 

Pulse sound alarm 

Pulse sound alarm 

Pulse sound alarm 

Pulse sound alarm 

Trigger remote 

alarm contact 

Trigger remote 

alarm contact 

Trigger remote 

alarm contact 

Trigger remote 

alarm contact 

Trigger remote 

alarm contact 

Trigger remote 

alarm contact 

Trigger remote 

alarm contact 

HXC-158/158B/358/358B/608/608B

Alarm

Summary of Contents for HXC-1308

Page 1: ...Technology Development Zone Qingdao 266510 P R China Web www haiermedical com Version 1st 2016 Dedicated code 0270501029 V13026 Read the Operation Manual carefully before using your appliance Keep the Operation Manual in a safe place Appearance color and layout of the door may vary Translation of the original instruction The scope of the product application For providing a storage environment of 2...

Page 2: ......

Page 3: ...rent from the figure in the manual due to product improvement The manual may be subject to changes without further notice Computer panel control digital display of temperature temperature display accuracy as 0 1 C and temperature range as 4 C 1 C temperature range of HXC 1308 and HXC 1308B as 4 C 1 5 c Multi fault alarm high and low temperature alarm low battery alarm power failure alarm sensor fa...

Page 4: ...fety Precautions Precautions for Use Product Installation Component Names Control Panel Application method Alarm Cleaning and Maintenance FAQ Refrigeration Schematic Circuit Diagram Specification 2 3 6 7 12 15 26 31 33 34 37 Content ...

Page 5: ...ct to prevent personal injuries and refrigerator damage In case of flammable gas leakage such as coal gas shut off the valve of leaked gas open windows for ventilation and exhaust do not plug in or pull out the power plug of refrigerator to prevent explosion and fire The refrigerator shall only be installed by technician or after sale maintenance staff Otherwise it may cause electric shock or fire...

Page 6: ...ugged into the socket to prevent fire Apply socket with ground wire to prevent electric shock If the socket fails to be grounded ground wires must be installed by technician Do not use the refrigerator outdoors Electric leakage or shock may be caused if the refrigerator gets wet by rainwater Do not place the refrigerator in humid places or places where the refrigerator may suffer splashing water T...

Page 7: ...to children And the flammable foaming needs to be disposed by professional persons CP IP foaming materials are flammable need professional processing HXC 608B 6 Drawers HXC 608 Blood Bank Refrigerator External dimensions W X D X H mm Internal dimensions W X D X H mm Effective volume Door Insulating materials Compressor Shelf Drawer Refrigeration mode Outer plate inner plate Condenser evaporator Re...

Page 8: ...e solvent such as gasoline Shut off the power as well as the battery switch if the refrigerator will be out of service for a long time 39 6 Precautions for Use Specification HXC 608 115V 60Hz Blood Bank Refrigerator External dimensions W X D X H mm Internal dimensions W X D X H mm Effective volume Door Insulating materials Compressor Shelf Drawer Refrigeration mode Outer plate inner plate Condense...

Page 9: ...W The refrigerator shall be grounded in a reliable manner If the socket is provided with grounding wire check if it is well grounded before operation If the socket is not provided with grounding wire a new one shall be installed by engineering technician Do not ground the refrigerator through gas pipes water supply pipes telephone lines or lightning conductors as electric shock may be caused The p...

Page 10: ...ailure alarm 48 hours when battery is fully charged I Temperature controller USB function temperature recorder Optional optional Y Electric shock resistance type Power connect style Climate Type Foaming Agent Noise Level 4 42dB A CP IP Note Climate Type 4 means the temperature is 30 the relative humidity is 55 Technical information may vary from the appliance that you have just purchased due to te...

Page 11: ... control panel Internal fan Display panel Sensor Defrosting Control Lower temperature Upper temperature Alarm VCC GND Heating strip at the cabinet outlet Heating strip of center sill Door heating transformer Door heating strip 1 Temperature recorder LED drive model Lamp switch USB drive panel USB port Reserved network interface Wiring terminal 2 Double throw switch Press wiring terminal box Power ...

Page 12: ...t the articles gradually into the refrigerator after complete examination of refrigerator 35 10 Circuit Diagram HXC 158 158B 358 358B 608 608B CN2 Cooling fan Compressor Defrosting heating strip CN2 CN1 CN3 CN11 BT1 BAT BAT Main control panel CN5 Internal fan Display panel CN2 CN3 Sensor Alarm Defrosting Control Lower temperature Upper temperature VCC GND Heating strip at the cabinet outlet CN1 t ...

Page 13: ...erature at one time If so the compressor may not stop for a long time with temperature reducing slowly which will reduce service life Articles must be put in batches and the temperature shall be reduced in steps until it reaches the temperature required Do not damage refrigerating loop Do not apply electronics without product permission in the refrigerator The refrigerator has memory function towa...

Page 14: ...mean data tested in the standard laboratory without noise in which the refrigerator without storage of articles is laid on the rubber pad and is steadily operated after closing the door to measure out the data at 1 meter from the surface the data do not include the value of startup and shutdown of the refrigerator During the process of usage it is normal that the actual noise is different from the...

Page 15: ... is received Consult us by phone for detailed information Purchasing of backup battery of the recorder Battery in the same specification and the same brand can be purchased in proper shops 9V DC In order to guarantee the reliability of the performance of the battery you can apply to our company for purchasing The ordering cycle is 1 2 months and it will be delivered 15 days after your money is rec...

Page 16: ...ipment and may lead to the occurrence of fault Compressor and other mechanical parts are in fully closed state and they do not need to be lubricated Clean the refrigerator Cleaning and Maintenance Component Cleaning In order to prevent from getting an electrical shock or personal injury make sure that the power of refrigerator is cut off before conducting any repair to the refrigerator Make sure t...

Page 17: ...battery switch is on or off the remote alarm can be started The remote alarm port coordinates with the audible alarm on the refrigerator Thus the remote alarm can be stopped by pressing the key of buzzer outage alarm cannot be canceled COM NO NC Temperature displayer The refrigerator is set to reach appropriate temperature automatically before leaving the factory 4 1 4 1 5 for HXC 1308 1308B Tempe...

Page 18: ...for the refrigerator He is in charge of examining the running status of the refrigerator and making the record If the temperature inside the refrigerator is too high or too low or other abnormal phenomenon he shall transfer the articles to other refrigerators which conform to the conditions till the fault has been settled and articles can be stored in that refrigerator again When the battery has f...

Page 19: ...back to the normal display mode The original temperature set value 4 flickers Please do the next operation within 60s Otherwise the display panel will return to display TS SN Alarm items Buzzing Light flashing 3 Fault of sensor Conditions of alarm start Display requirements Note Display board displays the alternation between the mean value of temperature of both upper and lower sensors and E3 wher...

Page 20: ...after a period of Fd user parameters can be set the default value is 30 min the buzzer and remote alarm will be restarted 1Hz Continuously Buzzing 1Hz Continuously Buzzing Alarm light flashes Display board displays the alternation between the temperature of upper sensor and E2 wherein E2 displaying for 2s while the temperature displaying for 6s HXC 1308 1308B Temperature recorder HXC 158 158B 358 ...

Page 21: ...ntil the pen returns to the paper At this moment the pen will pause at the most external temperature scaling line of the recording paper this scaling line does not have to be marked with temperature value but must be the one at the most external circle If the pen does not pause at the above position you can use 1 or 2 arrow key to adjust the position within 5 seconds to align the nib to the scalin...

Page 22: ...ng the change of temperature in the box In order to use the disk recorder in a right and reasonable way please read the instruction bellow carefully Get familiar with your temperature recorder Adjustment of Calibration of the Recorder The recorder is accurately adjusted before leaving the plant It will not be affected even if the power is cut off If adjustment is required please conduct it accordi...

Page 23: ...nction The computer panel has USB output function which can make use of U disk to export temperature and other test data Plug in U disk the machine begins to automatic identification and import data into U disk During data import press Conform calibration key the flashing display of USB on display panel represents the data export is being on and not yet complete After flashing for 5s it exits and ...

Page 24: ...r position can also be adjusted finely 3s later the scriber will slide to appointed scale according to the present perceived temperature then the calibration ends After retightening hold down bolt and pressing Key 3 again the recorder will switch to the process of paper calibration Please refer to the explanation of last page to calibrate the installation of paper After calibration the temperature...

Reviews: