background image

15

14

How to use the freezer

Unlocking the freezer

Be sure to unlock firstly for adjustment of set value.
1. Press the“       ” or “        ” and adjust to the number “06”.
2. Press  the“Set” key for 5 seconds when the “locked” lamp is OFF to get various settings under 
unlocking  status.
3. Press the “Set" key to set the temperature inside the box, high-temperature alarm and 
low-temperature alarm in a circular fashion while the corresponding indicator lamps are ON.

Setting the inner temperature

1. In unlocked mode, press “Set” key to select "Set Temp.", the temperature display flashes and 
displays the setting value.
2. Then, press“       ”or “      ” key, adjust the temperature setting value .
Temperature setting range:-10 to -86

. Recommended temperature setting range:-40 to -86

.

3. After adjustment, do not touch the unit for 10 seconds. The unit automatically enters the locked 
mode and the temperature display stops flashing which means the value set have been input into 
computer. Otherwise, the setting is invalid.
For example: Set the inner temperature to -80

.

4. After setting the inner temperature, the high temperature alarm and low temperature alarm will 
automatically adjust to proper values accordingly. 
If user has special requirements, follow the following steps to adjust the values manually.

Setting the high temperature alarm

1 .In unlocked mode, press “Set” key to select “High Temp. Alarm", the temperature display flashes 
and displays the setting value.
2. Then, press '      ' or '      ' key, adjust the high temperature alarm setting value.
Temperature setting range: at least +50

 above the inner temperature.

3. After adjusting, do not touch the unit for 10 seconds. The unit automatically enters the locked 
mode and the temperature display stops flashing which means the value set have been input into 
the computer. Otherwise the setting is invalid.
For example: If inner temperature is set to -80

, setting the high temperature alarm to -75

 is 

recommended.

Set

Control Panels

Set

V

Summary of Contents for DW-86L578S

Page 1: ... 再分批次放入物品 解锁 温度设定操作 按 或 键 显示板LED箱内温度显示区显示 00 按 调节至 06 默认值 可调 按住 键 5秒钟 解锁显示板 显示板锁定指示灯灭 再按一下 键 进入温度设定模式 在解锁状态下按下 设置 键 分别选择 箱内温度设定 高温报警设定 低温报警设定 相应指示灯亮了以后 按 或 键调整所需温度 修改完毕后 按 键保存箱内温度设定值 若20秒无按键 箱内温度设定值不保存 功能锁定后返回显示箱内温度 注意 其他内容请详见说明书 海尔生物医疗 智护生命科学 首次安装或搬动保存箱后 请顺时针旋转水平支脚使其支撑地面 确保保存箱不移动 安装固定 到位后需要静置24小时后再通电使用 为保证空气流通散热 保存箱四周至少保留30cm的间隙 保存箱使用环境温度为10 32 最佳环境温度为18 25 必要时应使用空调系统 湿度应低 于80 RH 环温32 时 湿度应低于57 R...

Page 2: ...选择 箱内温度设定 高温报警设定 低温报警设定 相应指示灯亮了以后 按 或 键调整所需温度 修改完毕后 按 键保存箱内温度设定值 若20秒无按键 箱内温度设定值不保存 功能锁定后返回显示箱内温度 注意 其他内容请详见说明书 海尔生物医疗 智护生命科学 首次安装或搬动保存箱后 请顺时针旋转水平支脚使其支撑地面 确保保存箱不移动 安装固定 到位后需要静置24小时后再通电使用 为保证空气流通散热 保存箱四周至少保留30cm的间隙 保存箱使用环境温度为10 32 最佳环境温度为18 25 必要时应使用空调系统 湿度应低 于80 RH 环温32 时 湿度应低于57 RH 每台保存箱应该独立使用一个插座 且电源插座承受电流应大于16A 并可靠接地 通电时先打开保存箱电控箱上的电池开关 需长时间关机时 先拔掉电源再关掉此开关 保存箱每次开门时间尽量不超过1分钟 关门前如门封条上有冰水 建议擦拭干净 以保...

Page 3: ...products your Haier ULT freezer may be different from the ones illustrated in this manual and we do apologize for this User manuals are subject to change without notice This product is suitable for the ultra low temperature storage of products in applications such as clinical pharmaceutical scientific research and epidemic institutions It also can be used in blood stations hospitals centers for di...

Page 4: ...IOT board Optional screen Tricolor light IOT board Backup system control board Backup system sensor Optional touch screen Solenoid valve Electromagnet lock Electromagnet lock Backup system board Communication port for display board Magnetic proximity door switch Door ajar switch Cold chain Printer Tricolor light Optional Lithium battery Acid lead battery Battery switch Battery switch Battery Batte...

Page 5: ...the descriptions marked with the following labels Safety labels Haier Biomedical UK LTD Address Mytogen House 11 Browning Road Heathfield TN21 8DB United Kingdom Symbol for Manufacture R600a R1150 R600a R1150 Install padlock here for best security Wear cryo gloves to prevent frostbite 3 Refrigeration and Wiring Diagram Refrigeration diagram Suction Accumlator Capillary Tube Cascade Condenser Evapo...

Page 6: ... other methods should be considered in order to ensure safe working environments If there is a leakage of petroleum gas or other flammable gas close the gas supply valve and open doors and windows to ventilate the air Do not plug or unplug your freezer unit in order to avoid potential explosion or fire Only professional technicians or Haier service personnel can install the unit Failure to do so m...

Page 7: ...lug thereby causing electric shock and fire Should there be any malfunction in the equipment power off the unit and unplug the power cord from the power supply Continuous operation in an abnormal condition may result in electric shock and fire Never use gas lines water mains telephone lines or lightening rods as the grounding device for your Haier ULT unit This type of improper grounding may cause...

Page 8: ... on top of the unit or place any object on it Falling equipment may cause injury or property damage Do not use electrical appliances inside the chamber of the appliance unless they are of the type recommended by the manufacturer Optional Accessories When using the temperature recorder please refer to the User manual for Temperature Recorder provided with the recorder Temperature recorder The tempe...

Page 9: ...ration system During any repair operations gloves should be worn to prevent getting injured by sharp edges or corners Hold firmly onto the handle to close the door to avoid pinching your hands Do not tilt the unit more than 45 degrees when moving the unit The air filter for the condenser should be cleaned regularly A dirty filter could cause a malfunction or the freezer temperature to rise Do not ...

Page 10: ... the paint coating and plastic components Particularly never use gasoline or a solution with volatile chemicals to clean plastic or rubber parts Defrost the interior Frost and ice can form in between the door gasket and frame to form an air gap which can decrease the refrigeration effect of the unit Please use the provided plastic scraper to defrost the interior doors The following steps are how t...

Page 11: ...on before using the equipment If the socket is not grounded it must be connected to a grounded terminal by a qualified technician Installation Installation environment Installation site Do not ground the freezer through gas lines water mains telephone lines and lightning rods as this may lead to electric shock After installation the power plug must be easy to reach In case of emergency it is easy ...

Page 12: ...the battery sustains the temperature display If the battery voltage is insufficient the temperature display will turn off While the battery is still capable of providing power to the display the temperature display can be turned off by unplugging the power cord and turning off the battery control switch The freezer is also designed to auto adjust the inner temperature set point in high ambient tem...

Page 13: ...etwork system is in service Running indicator lamp is ON It means that the low temperature compressor is in service Backup system indicator lamp is ON It means that the backup refrigeration system is in service Alarm Alarm Status Code Instruction Buzzer Remarks High temperature alarm The current temperature the set one for high temperature alarm for 1 minute on end E00 The alarm lamp flashes ON an...

Page 14: ... recovery the unit can resume its operation based on the previous settings In the event of a power outage and recovery be sure to wait for at least 5 minutes before turning the unit on again to avoid damaging the compressors and refrigeration system If the freezer is not in use in an area without any supervision please make sure children will not approach the freezer and the doors should not be cl...

Page 15: ...r key to modify the set value of usb data reading cycle 3 T2 usb timing MM P2 01 12 DD P3 01 31 YYYY Pl 10 99 Hour P4 00 23 Minute P5 00 59 Press the or key to modify the timing of usb 4 P6 12 is the default value for USB derivative mode it is adjustable from 0 to 12 The data can be exported in different time frames with parameters of 0 12 among which 12 as default means exporting the data generat...

Page 16: ...play stops flashing which means the value set have been input into computer Otherwise the setting is invalid For example Set the inner temperature to 80 4 After setting the inner temperature the high temperature alarm and low temperature alarm will automatically adjust to proper values accordingly If user has special requirements follow the following steps to adjust the values manually Setting the...

Reviews: