ULT Energy-saving Series
After-sales Service Manual
42
A
-----------Condenser
B
-----------Dry filter
C
-----------One stage compressor
D
-----------Capillary tube
E
----------- Gas-liquid separator
F
-----------Heat exchanger
G
----------- Pressure switch
H
-----------Oil separator
I
----------- Two stage compressor
J
----------- Evaporator
XI. Installation and disassembly process
11.1 Installation location:
11.1.1 Avoid exposure to direct sunlight
11.1.2 The surrounding air is well-ventilated
11.1.3 Avoid excessive dust
11.1.4 Avoid mechanical swing or vibration
11.1.5 The ambient temperature is between 10°C and 28°C; the maximum temperature is not more than 32°C. The
optimal temperature is between 18°C and 25°C, and the air conditioning system shall be used if necessary.
11.1.6 Height of equipment working position: lower than 2000 m
11.1.7 Working humidity: lower than 80% RH If the maximum working temperature is 32°C, the humidity should be
lower than 57% RH.
11.1.8 Input voltage: Within 220±10%.
*Note: Because the ULT Freezer is sensitive to the ambient temperature, if installed in an environment other than
the above, it will not operate normally. Please use the freezer after improving the environment.
11.2 Product handling and package removal
11.2.1 Handling: The product is very heavy, so forklift or pallet jack shall be used to avoid damage to personnel and
B
D
F
E
G
H
J
A
I
C
B