10-6
Shroud and Base Assemblies
P/N 108755-US, Manual revision: AA
B
AG
P
IN
I
NSTALLATION
1.
Align one pin over each hole from the outside of the shroud. Depending
on the hardware ordered or provided with the Shaker, place the backing
plate or ¼ in. -20 x 0.38 in. pan head phillips screw through the washer
or plate and shroud into the pin. Partially tighten the screw into the pin.
Repeat this process for the remaining pin.
2.
To obtain the correct pin spacing, position a Haemonetics certified cali-
bration weight over the pins and adjust the spacing while tightening the
screws into the pins. When the proper spacing has been obtained, secure-
ly tighten the screws in place.
Summary of Contents for SEBRA 1040
Page 1: ...108755 US service manual Model 1040 Digital Blood Shaker...
Page 3: ......
Page 4: ...Printed in USA Copyright 1999 2010 Haemonetics Corporation...
Page 8: ...vi Preface P N 108755 US Manual revision AA...
Page 12: ...x Table of Contents P N 108755 US Manual revision AA...
Page 14: ...xii List of Figures P N 108755 US Manual revision AA...
Page 20: ...1 6 Introduction P N 108755 US Manual revision AA...
Page 36: ...3 10 Inspection Procedures P N 108755 US Manual revision AA...
Page 40: ...4 4 Parts Access P N 108755 US Manual revision AA...
Page 48: ...5 8 Strain Gauge P N 108755 US Manual revision AA...
Page 60: ...7 6 Tube Clamp Assembly P N 108755 US Manual revision AA...
Page 64: ...8 4 PCB Chassis Assembly Installation P N 108755 US Manual revision AA...
Page 86: ...11 4 Clamp Body Boss Replacement P N 108755 US Manual revision AA...
Page 98: ...13 8 Cap Assembly P N 108755 US Manual revision AA...
Page 102: ...14 4 Power Supply P N 108755 US Manual revision AA...
Page 105: ......