Microprocessor
9-3
P/N 108755-US, Manual revision: AA
P
ROGRAM
SETTING
"F1"
1.
Ensure that the Shaker is powered "on." Press <#> then <1>. The display
will read "pw." Enter default password by pressing <-> six times.
2.
Press <0>, then <*> thirteen times. The display will read "f1: 02."
3.
Press <+> twice. The display will read "f1: 04." Program step "f1" is now
set.