Troubleshooting during an MCS+ Procedure
7-7
P/N 85213-30, Manual revision: B
AC depletion
procedure
During an MCS+ collection procedure, the anticoagulant solution may be
depleted before the target collection product volume has been reached. If the AC
solution bag is empty, the operator will hear an alarm and receive the screen
message “AIR DETECTED BY THE ACAD”.
Warning: Haemonetics recommends that the collection of blood components
be discontinued at this point.
The operator can attempt to return any blood components remaining in the bowl
using manual gravity infusion.
Summary of Contents for LN9000-110-EWC
Page 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Page 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Page 20: ......
Page 30: ......
Page 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Page 85: ......
Page 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Page 100: ......