8-8
Consulting Reference Information
P/N 85213-30, Manual revision: B
IEC 60601-1-2, Table 202: Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment guidance
*Electrostatic dis-
charge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least
30%.
*Electrical fast
transient burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 2 kV for power
supply lines
Mains power quality should be that of a typi-
cal commercial or hospital environment.
*Surge
IEC 61000-4-5
± 1 kV differen-
tial mode
± 2 kV common
mode
± 1 kV differen-
tial mode
± 2 kV common
mode
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Voltage dips,
short interrup-
tions, voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95%
dip in U
T
) for 0.5
cycles
40% U
T
(60%
dip in U
T
)
for 5 cycles
70% U
T
(30%
dip in U
T
)
for 25 cycles
< 5% U
T
(> 95%
dip in U
T
) for 5
sec
< 5% U
T
(> 95%
dip in U
T
) for 0.5
cycles
40% U
T
(60%
dip in U
T
)
for 5 cycles
70% U
T
(30%
dip in U
T
)
for 25 cycles
< 5% U
T
(> 95%
dip in U
T
) for 5
sec
Mains power quality should be that of a typi-
cal commercial or hospital environment.
If the operator of the Haemonetics system
requires continued operation during power
mains interruptions, it is recommended that
the Haemonetics system be powered from an
uninterruptible power supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.
* Authorized maximum value
Summary of Contents for LN9000-110-EWC
Page 2: ...2002 2006 Haemonetics Corporation All rights reserved...
Page 10: ...x Preface P N 85213 30 Manual revision B The Haemonetics MCS LN09000 220E ED...
Page 20: ......
Page 30: ......
Page 81: ...Maintaining the MCS Equipment 5 7 P N 85213 30 Manual revision B...
Page 85: ......
Page 86: ...5 12 Maintaining the MCS Equipment P N 85213 30 Manual revision B...
Page 100: ......