The menu is used via the four buttons.
- It is opened with the 

MENU

 button.

- You can scroll through the menu options / submenu options with the 

 buttons, or increase or decrease 

adjustable values.
- The 

enter

 button is for opening the selected menu option or to confirm the entry.

- The menu is closed with the 

escape

 button. The device is then in measuring mode.

4.2.1. „About“ menu option

With this menu option, you can read out the following information about your device:

Firmware 

Version of the installed software

Data last cal.  Date of the last calibration
Serial No. 

Serial number

Procedure

:

Press 

MENU

 to open the menu

Open menu item “1. About” with 

enter

Select the desired submenu option with

 

 and open it with 

enter

 

The device-specific information can be read in the display
Return to the submenu with 

enter

Select another submenu option with

 

or
Select the submenu option “1.4 Back” with 

 and return to the main menu with 

enter

or

Exit the menu with 

escape

. The device returns to measuring mode.

4.2.2. „Configuration“ menu option

With this menu option, you can configure the following settings:
Auto off

Select the time after which the device automatically switches off to save batteries.

You can select one of the following: off, 60, 30, 8 min. (Factory setting: 60 min)

Averaging

Here, you can select the desired measuring rate customized for your application, at the 
same time, the frequency also changes with which the measured value display is up-
dated in the display. The lower measuring rate is suited to slower loads, the higher rate 
is suited to short, fast load changes. The higher the measuring rate, the better the detec-
tion of peak loads. However, with the high measuring rate the measured value display 
becomes slightly more disturbed, as the noise element increases.
You can select one of the following: 

Settings

Measuring 
rate

Display frequency

slow

10 Hz

2 Hz

middle

20 Hz

5 Hz (factory setting)

fast

100 Hz

10 Hz

Summary of Contents for PAD2

Page 1: ... Description 2 1 2 Scope of supply 2 1 3 Intended use 2 1 4 Safety instructions 2 2 Technical Data 3 3 Operating Elements and Functions 4 4 Operation 5 4 1 Switching on 5 4 2 Using the menu MENU 5 4 2 1 About menu option 6 4 2 2 Configuration menu option 6 4 3 Measuring 7 4 3 1 Zero adjustment and taring 0 T 8 4 3 2 Switching the measured value display PEAK 8 4 3 3 Performing the measurement 8 4 4...

Page 2: ... its nominal load The nominal load can be found on the type label of the sensor You can also find this infor mation on the force gauge display when switching it on The force gauge must not be subjected to any forces This force gauge is intended only for the exclusive use in areas that are not subject to calibration requirements It is a highly sensitive measuring instrument and must be treated acco...

Page 3: ...orce gauges display when switching on Display 12 digits 2 lines Resolution 16 bit Power supply of PAD2 4 x battery type AA Operating time 8 h continuous operation Power supply of force sensor 3 3V Nominal resistance of strain gauge 100Ω to 500kΩ Auto off time 8 30 60 min off Operating temperature range From 0 to 50 C 32 to 122 F Storage temperature range From 20 to 70 C 4 to 158 F Protection class...

Page 4: ...n this mode the force gauge is calibrated in the factory Button Name in the measuring mode Function in the measuring mode grey font Name in the menu Function in the menu light blue font ZERO POSITIONING TARING Zero adjustment taring and Clear peak values up scroll up in the menu increase numeric values PEAK Toggle between displays of the actual measured value of the max value largest value in posi...

Page 5: ...ake sure that this load is not exceed ed With the connected sensor the current measured value is shown in the display Please note that a previously tared value has been retained The peak values were deleted 4 2 Using the menu MENU The menu provides the following options Read out information about the device About menu option Configure device settings Configuration menu option Calibrate the device ...

Page 6: ...be read in the display Return to the submenu with enter Select another submenu option with or Select the submenu option 1 4 Back with and return to the main menu with enter or Exit the menu with escape The device returns to measuring mode 4 2 2 Configuration menu option With this menu option you can configure the following settings Auto off Select the time after which the device automatically swit...

Page 7: ...afety regulations The force gauge or other material assets may be damaged To adapt the PAD2 to existing lifting devices only lifting gear swivel heads eyebolts etc may be used that correspond at least to the nominal force of the sensor The operator is responsible for selection and control Product liability and warranty of the manufacturer exclusively refers to the standard equipment of the force g...

Page 8: ...xiting the menu the current measured value is always shown in the display You can also select display of the peak values with the PEAK button In this setting the measured value display corresponds to a drag indicator which is further advanced in case of increasing decreasing values If you press the PEAK button once then the maximum value is displayed in positive load direction Peak Max is displaye...

Page 9: ...teries polarity in doing so Slide the lid along the lateral guide back onto the battery compartment and let the nose click into place in the housing by applying slight pressure Thread the housing first on the cable side and then on the opposite side back into the protective cover 4 5 Turning off By holding the button down the force gauge is switched off The peak values are deleted If the system wa...

Page 10: ...NE Elektronische Messgeräte GmbH Heinrich Hertz Str 29 D 40699 Erkrath Germany Telefon 0211 9 2591 0 Fax 0211 9 2591 20 http www haehne de Email info haehne de PAD2 Technical modifications reserved Page 10 ...

Page 11: ... auto off averaging measurement rate setting and Contrast These are described in the Operating Manual PAD2 The third menu item contains the factory settings calibration This area is password protected with a 4 digit number During each calibration a counter is counted up irrevocably this count reading is displayed in each case before the re calibration The settings in the service menu 3 must be con...

Page 12: ...ounter If canceled with Esc the counter is not increased Note Termination is currently available only at the display Load is still changed Load Here the applied test load can be specified in 10 increments of the rated load Pkt 1 min The zero point is set without load Pkt 2 max Max is set with X of load The calibration process is completed successfully for 1 5 seconds Cal Ready appears on the displ...

Page 13: ... level 3 Residential area Interference field strength Industrial sector Electrostatic discharges ESD Electromagnetic fields Electromagnetic fields mobile phone radiation Before commissioning the PAD2 it is necessary to ensure that the system consisting of the portable strain gauge force gauge PAD2 and a connected sensor is according to the EC Guidelines EMC in case of modifications to the instrume...

Page 14: ...HAEHNE Elektronische Messgeräte GmbH Heinrich Hertz Str 29 D 40699 Erkrath Germany Telefon 0211 92591 0 Fax 0211 92591 20 http www haehne de Email info haehne de HAEHNE ...

Reviews: