3. Service
The following settings can be configured here.
To save the set values must be confirmed with the key „Enter“ until the menu item has been left
The „Service“ menu is protected by a 4-digit password (currently the last 4 digits of the serial number which is
located on the rear panel)
While entering the password use the „Esc“ key to navigate one point back if a number was incorrectly entered
and confirmed. Press „Enter“ to confirm the single digits.
3.1 Calibration
The following operations are performed:
(each point is confirmed with „Enter“)
Cal. count
Display of the number of previous calibrations.
Only fully traversed calibrations lead to the increase of the counter.
If canceled with „Esc“, the counter is not increased
Note: Termination is currently available only at the display „Load [%]“ - is still changed!
Load [%]
Here, the applied test load can be specified in 10% increments of the rated load.
Pkt. 1 (min.)
The zero point is set (without load)
Pkt. 2 (max.)
Max is set (with X% of load)
The calibration process is completed successfully, for 1.5 seconds „Cal Ready“ appears
on the display, the calibration values are stored. It goes to 3.2 immediately.
3.2 Settings
The following operations are performed
(each point is confirmed with „Enter“)
Nominal Load
Enter the amount of the nominal load (max. 99999).
While entering use „Esc“ key to navigate back one digit if a number was
incorrectly entered and confirmed. Press „Enter“ to confirm the single digits.
Dez. Pos
Here can be specified the number of decimal places after the decimal point.
(0,0 / 0,00 / 0,000)
Round
Indication of the rounding of values of the last displayed digit
(1 / 2 / 5 / 10 / 20 / 50 / 100)
Unit
The unit can be given (N, daN, kN, g, kg, t, lb)
3.3 to 3.6 for factory internal purposes
3.7 Back