Troubleshooting 

Handpiece  vibrates,  is  very  noisy 

and/  or  becomes  warm  in  the 
front part  

Use  of  bent  burs/  tools 

  Use  a 

different  bur/  tool  to  test  this.  The 
bur/tool  itself  can be  tested  with  the 
hadewe  Bur  Shaft  Tester  (Art.  No. 
(4990) to see if it is bent). 

 
 

Maximum  allowed  speed  of  the  tool 
was  exceeded,  which  caused  a 
stronger vibration 

 pay attention to 

the  tool  (i.g.  bur,  capper  and 
instrument)  manufacturers  handling 
details. 

 
Bur/tool  can  be  inserted  only  with 

difficulty, or not at all.

 

Use  of  bent  burs/  tools 

  Use  a 

different  bur/  tool  to  test  this.  The 
bur/tool  itself  can be  tested  with  the 
hadewe  Bur  Shaft  Tester  (Art.  No. 
4990) to see, if it is bent). 

 

Chuck  is  dirty 

  Clean  the  chuck 

according  to  the  instructions  in  the 
manual. 

 

Bur does not clamp, or does not spin 

Chuck  is  not  tightened  well 

 

Tighten  the  chuck  firmly  according 
to the instruction in the manual. 

 

Knob  of  the  handpiece  can  not  be 

moved easily 

 

Chuck  is  dirty 

  Clean  the  chuck 

according  to  the  instructions  in  the 
manual. 

 
 

Unit does not work at all. 

 
 

Pull  out  the  mains  plug  to  reset  the 
device 

 

The  unit  shows  different  sound 

levels at different sites of use 

 

The  sound  level  varies  according  to 
the  surroundings  of  the  unit.  If  you 
set  up  the  unit  directly  in  front  of  a 
wall, the sound is reflected stronger, 
as  if  it  would,  while  standing  further 
away from a wall. 

 

Handpiece  motor  stutters  shortly 

(appr. 1 sec) when turning unit on 

 

In  very  rare  cases  this  effect  did 
appear.  It  will  not  harm  or  damage 
the device. 

 
 

Summary of Contents for Xantos

Page 1: ...Manual Xantos Art No 0897 0898 FB 04 280 Rev 1 Art no 0897 0898...

Page 2: ...tions 3 Cleaning Disinfection 7 Getting to know the unit 8 Before the use 9 Area of application 9 Installation site location 9 Maintenance Care 11 Cleaning the chuck 11 Handling details 12 Troubleshoo...

Page 3: ...extracted air can escape well Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely in your hand To avoid the risk of electrocution only connect the d...

Page 4: ...tion or spray technique Wear a nose and mouth protection Always follow the safety instructions provided by the manufacturer of the materials being formed and used For safety reasons wear goggles durin...

Page 5: ...Before using the equipment assure that the tool can be pushed far enough into the chuck Test this with a dipstick and clean the handpiece if necessary the inserted tool sits properly in the chuck and...

Page 6: ...ly to be accomplished by an authorized specialist The mains cable may also only be changed by the manufacturer or an authorized specialist Never remove the housing especially not while the equipment i...

Page 7: ...capper etc first Then handle the cleaning of the tool according to the instructions of it s manufacturer Remove tissue remains by wiping off the unit with a mildly moistened paper cloth Afterwards cle...

Page 8: ...8 Getting to know the unit 1 Speed display 2 LED shines during left rotation 3 Increase speed 4 change directory 5 decrease speed 6 on off button 8...

Page 9: ...ental technique laboratory Milling trimming and polishing work at plaster casts and models Do not use this unit for dentist purposes i e working directly on the patient Industry Grinding milling and p...

Page 10: ...th a diameter of 2 35 mm this is the diameter usually used for tools in podiatry To open the chuck push the spanner forward with your thumb and leave it there Now insert the tool as far as possible Pu...

Page 11: ...To assemble the chuck follow the steps in reverse order Make sure that the chuck is opened during the assembly the handpiece button must be in forward position and is firmly applied afterwards Cleani...

Page 12: ...hen you not working with the handpiece always remove the bur tool from it Always make sure that you do not let the handpiece fall down or exposes it to any shocks The ball bearings and the motor could...

Page 13: ...if it is bent Chuck is dirty Clean the chuck according to the instructions in the manual Bur does not clamp or does not spin Chuck is not tightened well Tighten the chuck firmly according to the inst...

Page 14: ...relates are conform to the following standards or normative documents 2014 30 EU directive EMC 2014 35 EU low voltage directive hadewe GmbH Grambartstra e 10 30165 Hannover Germany Symbols Caution Im...

Page 15: ......

Page 16: ...hadewe GmbH Grambartstra e 10 30165 Hannover Germany info hadewe de www hadewe de...

Reviews: