background image

 

 

10 

Maintenance & Care 

Starting  up  and  operating 
the unit: 

Put  down  the  handpiece  in  the 
handpieceholder.  Plug-in  the  plug. 
Turn on the unit with button 6. With 
button  3  and  5  you  can  adjust  the 
speed  level.  Regulary  the  drill  runs 
clockwise.  To  change  the  dircetion 
push  button  4.  While  the  drill  runs 
with  the  left rotation the green LED 
(no 2 in pic.1) is shining.

 

 

 
Clamping the tool/ bur 

The  handpiece  is  equipped  with  a 
chuck,  which  can  hold  all  burs  with 
a  diameter  of  2,35  mm  (this  is  the 
diameter  usually  used  for  tools  in 
podiatry).  To  open  the  chuck,  push 
the  spanner  forward  with  your 
thumb and leave it there. Now insert 
the tool as far as possible. Pull back 
the  knob  and  the  tool  will  be  firmly 
clamped. 
 

 

 

 

 

For 

security 

reasons, 

only  change  the  tool,  if 
the  handpiece  is  turned 
off. 

 

 

 

 

Setting up the speed 

Hold  the  handpiece  (incl.  tool)  in 
your  hand  or  store  it  in  the 
handpiece  holder  attached  to  the 
side of the control unit. 
 
Switch  on  the  unit  with  button  (6). 
Now  you  can  also  set  the  speed 
with buttons (3) and (4), and choose 
its  direction  of  rotation  with  button 
(5).  In  reverse  running  the  LED  (2) 
will be on.  

Summary of Contents for Xantos

Page 1: ...Manual Xantos Art No 0897 0898 FB 04 280 Rev 1 Art no 0897 0898...

Page 2: ...tions 3 Cleaning Disinfection 7 Getting to know the unit 8 Before the use 9 Area of application 9 Installation site location 9 Maintenance Care 11 Cleaning the chuck 11 Handling details 12 Troubleshoo...

Page 3: ...extracted air can escape well Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely in your hand To avoid the risk of electrocution only connect the d...

Page 4: ...tion or spray technique Wear a nose and mouth protection Always follow the safety instructions provided by the manufacturer of the materials being formed and used For safety reasons wear goggles durin...

Page 5: ...Before using the equipment assure that the tool can be pushed far enough into the chuck Test this with a dipstick and clean the handpiece if necessary the inserted tool sits properly in the chuck and...

Page 6: ...ly to be accomplished by an authorized specialist The mains cable may also only be changed by the manufacturer or an authorized specialist Never remove the housing especially not while the equipment i...

Page 7: ...capper etc first Then handle the cleaning of the tool according to the instructions of it s manufacturer Remove tissue remains by wiping off the unit with a mildly moistened paper cloth Afterwards cle...

Page 8: ...8 Getting to know the unit 1 Speed display 2 LED shines during left rotation 3 Increase speed 4 change directory 5 decrease speed 6 on off button 8...

Page 9: ...ental technique laboratory Milling trimming and polishing work at plaster casts and models Do not use this unit for dentist purposes i e working directly on the patient Industry Grinding milling and p...

Page 10: ...th a diameter of 2 35 mm this is the diameter usually used for tools in podiatry To open the chuck push the spanner forward with your thumb and leave it there Now insert the tool as far as possible Pu...

Page 11: ...To assemble the chuck follow the steps in reverse order Make sure that the chuck is opened during the assembly the handpiece button must be in forward position and is firmly applied afterwards Cleani...

Page 12: ...hen you not working with the handpiece always remove the bur tool from it Always make sure that you do not let the handpiece fall down or exposes it to any shocks The ball bearings and the motor could...

Page 13: ...if it is bent Chuck is dirty Clean the chuck according to the instructions in the manual Bur does not clamp or does not spin Chuck is not tightened well Tighten the chuck firmly according to the inst...

Page 14: ...relates are conform to the following standards or normative documents 2014 30 EU directive EMC 2014 35 EU low voltage directive hadewe GmbH Grambartstra e 10 30165 Hannover Germany Symbols Caution Im...

Page 15: ......

Page 16: ...hadewe GmbH Grambartstra e 10 30165 Hannover Germany info hadewe de www hadewe de...