Troubleshooting 

Handpiece vibrates, is very noisy 
and/or becomes warm in the front 
part.  

Use of bent burs/tools. 

 Use a 

different bur/tool to test this. The 
bur/tool  itself  can  be  tested  with 
the Bur Shaft Tester (art. 4990) to 
see if it is bent. 

 

 

Maximum  allowed  speed  of  the 
tool 

was 

exceeded, 

which 

caused  a  stronger  vibration. 

 

Pay attention to the tool (e.g. bur, 
capper, 

and 

instrument) 

manufacturers handling details. 

 

Bur/tool can be inserted only with 
difficulty or not at all.

 

Use of bent burs/tools. 

 Use a 

different bur/tool to test this. The 
bur/tool  itself  can  be  tested  with 
the Bur Shaft Tester (art. 4990) to 
see if it is bent. 

 

It does not spray. 

Pressure cannot be set up in the 
tank,  because  the  lid  is  not 
tightened well.  

 

There  are  parts  of  dirt  in  the 
spray system. 

 Clean the spray 

system 

according 

to 

the 

instructions in the manual. 

 

The filter (art. 3497) in the tank is 
blocked. 

 Change the filter.  

 

Unit does not work at all. 

Check  if  the  main  switch  of  the 
unit is turned on at the side of the 
unit. 

 

 

Foot switch is used constantly or 
faulty. 

  Pull  out  the  socket  of 

the foot switch to see if this is the 
cause. 

 

Pull  out  the  mains  plug  to  reset 
the device. 

 

After  lengthy  use  unit  does  not 
spray.

 

 

Unit is overheated. 

 Slots of the 

unit  should  not  be  covered  or 
blocked.  Let  cool  down  the  unit 
for a while. 

 

The  unit  shows  different  sound 
levels at different sites of use. 
 

The sound level varies according 
to the surroundings of the unit. If 
you set up the unit directly in front 
of  a  wall,  the  sound  is  reflected 
stronger,  as  if  it  would,  while 
standing  further  away  from  a 
wall.   

 

The  motor  of  the  handpiece 
stutters 

(e.g. 

sec.) 

while 

switching it on. 

 

Very  seldom  this  error  might 
occur.  It  would  not  damage  the 
motor at all. 

 

Spray stream stutters. 

 

There is air in the spray system. 
After 5 min running, this effect will 
disappear. 

 

Acceleration of the spray stream. 

Note: To reduce the starting time 
of  the  spray,  turn  the  spray 
regulator on Max. After the spray 
started,  adjust  the  stream  to  the 
requested amount of water. 

 

Fluid leaks out of the unit. 

Please be aware of the maximum 
filling  level  of  the  tank.  Do  not 
exceed the maximum level when 
filling the tank. 

Summary of Contents for HP-S2

Page 1: ...Article 0592 FB 04 448 Rev 1 Date 28 11 2023 Manual HP S Automatic...

Page 2: ...11 Startup Procedure 12 Application 13 Work Steps 13 Work Positioning 14 Spray Liquid 16 Insert Bur 17 Adjusting the Speed 17 Memory Buttons 17 Spray Function 17 Drying Function Tool Change 18 Foot S...

Page 3: ...t Keep it somewhere safe If you give this product to someone else please include this document as it includes important information on how to use the product correctly Short description of functions T...

Page 4: ...at the power plug is always accessible The power switch should also always be easily accessible Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely...

Page 5: ...a face mask conforming to EN 14683 type IIR or EN 149 FFP 2 must also be worn Ensure that the patient s foot and the handpiece are held firmly in place during the work To prevent injury work with extr...

Page 6: ...ol sits firmly in the chuck do not work with tools which have an oily shaft are worn are bent Before using the equipment assure that the tool can be pushed far enough into the chuck test this with a d...

Page 7: ...rry out any repairs modifications or maintenance work yourself This is only to be accomplished by an authorized specialist The mains cable may also only be changed by the manufacturer or an authorized...

Page 8: ...leaning and user maintenance shall not be made by children without supervision This appliance is not intended for use by persons including children with reduced physical sensory or mental capabilities...

Page 9: ...here to the cleaning and disinfection instructions provided below After every treatment please carry out these steps Remove the rotating tool e g bur first To do so pay attention to the instructions o...

Page 10: ...F ON spray 3 decrease speed 4 increase speed 5 changeover switch for clockwise or anticlockwise operation 6 LED is ON during reverse rotation 7 memory button for rotation speed M1 M2 8 Start Stop butt...

Page 11: ...or can be used as a portable unit e g in a carry case It is suitable for the following scopes foot care milling trimming and polishing works manicure grinding milling and polishing works The devise is...

Page 12: ...oor Image Example of positioning of foot care unit in treatment room Image Bird s eye view of work area If you have a foot switch insert its plug into the port on the right of the device which display...

Page 13: ...for clockwise use The rotation direction is irrelevant for diamond cutters and high grade corundum sanders in terms of efficiency Working anti clockwise can be beneficial if for example the dust obscu...

Page 14: ...Application Work Positioning The following images are examples of how to work on various parts of the foot Always make sure that the foot is firmly gripped and that you can work freely and without str...

Page 15: ...15 Application...

Page 16: ...the housing Switch on the power switch in the rear of the unit Never use tap water Danger of calcification Also do not use explosive or combustible liquids e g alcohol Note Never refill the tank dire...

Page 17: ...Frequently used speed levels can be saved Therefore three memory buttons are available At first choose your speed level and then save it by pushing button 7 m1 for 4 seconds To recall the above speed...

Page 18: ...his function can also be used to change your tools while the handpiece is not running Foot Switch Connecting with control unit Only use the original foot switch art 0970 to avoid damages Insert the pl...

Page 19: ...m Push the new washer in the top Please pay attention to the way it was installed Screw the top onto the handpiece tip Filter Even smallest parts of dirt could block the nozzle which sits in the top A...

Page 20: ...of the whole spray system Parts of dirt are rinsed into the tank Do not use this liquid again Clean the tank immediately after realizing a blockage in the nozzle Otherwise the dirt particles might sti...

Page 21: ...disinfection liquids This extraneous rost affects internal parts of the handpiece in form of rust bloom and causes pitting Therefore when you not working with the handpiece always remove the bur tool...

Page 22: ...d on at the side of the unit Foot switch is used constantly or faulty Pull out the socket of the foot switch to see if this is the cause Pull out the mains plug to reset the device After lengthy use u...

Page 23: ...N EN ISO 1797 1 Operating Time The device is designed for extended usage Ambient Conditions The device is intended for use in dry closed rooms Operation Temp 10 C to 26 C Rel humidity 25 to 75 not con...

Page 24: ...rderly disposal of your unit please contact your distributor in your country Safety Check A safety check must be done in compliance with the regulations of the country where the device is used The ins...

Page 25: ...s the lower and upper temperature limit air pressure permissible range air humidity permissible range Keep dry alternating current direct current safety fuse type B applied part socket foot switch dat...

Page 26: ......

Page 27: ......

Page 28: ...hadewe GmbH Grambartstra e 10 30165 Hannover Germany info hadewe de www hadewe de...

Reviews: