15
4-4. Doppler Arterial Testing
Doppler Arterial Testing includes ABI, TBI and PV-Arterial studies for up to 6 sites
with semi-automatic simple operation. PPG and Doppler probes are used for taking
blood pressures on great toes, ankles and Arms, respectively.
4-4-1. Preparations
(1) Get the necessary components ready as the following example for 6 sites:
Cuffs
(Option)
VC-10: 4 pcs.
For brachia and ankles
DVC-1.9: 2 pcs.
For great toes
Tubing
2 tubing each color x 3 sets;
White for Arm, Green for Ankle/PT, Black for Great toe
(Tubing with blue mark at both ends are for the right side)
PPG probe
PG-01 x 2 pcs.
PPG clip
PPG Clip x 2 pcs.
Doppler probe
(Option)
BT8M05S8C x 1pc.
Ultrasonic gel (TSUBAKIYA)
for Doppler testing (Option)
(2) Click
Doppler Arterial Testing
launcher icon on Main Screen.
(3)
Settings
Click
Settings
button on Doppler Arterial Testing screen and set Doppler-test
settings as follows.
<Doppler Arterial Testing screen>