9
4-2. Automatic Arterial Testing (Oscillometry)
Automatic Arterial Testing includes ABI, TBI and PV-Arterial studies for up to 6 sites
with simple one-button operation. PPG and oscillometry are used for taking blood
pressures on great toes, ankles and Arms, respectively.
4-2-1. Preparations
(1) Get the necessary components ready as the following example for 6 sites:
Cuffs
(Option)
VC-10: 4 pcs.
For brachia and ankles
DVC-1.9: 2 pcs.
For great toes
Tubing
2 tubing each color x 3 sets;
White for Arm, Green for Ankle/PT, Black for Great toe
(Tubing with blue mark at both ends are for the right side)
PPG probe
PG-01 x 2 pcs.
PPG clip
PPG Clip x 2 pcs.
(2) Click
Automatic Arterial Testing
launcher icon on Main Screen.
(3)
Settings
Click
Settings
button on Automatic Arterial Testing screen and set the
Auto-testing order as follows.
Auto-testing order:
No.1: Arm (R/L) - Great Toe (R/L) - Ankle (R/L)
No.2: Arm (R/L) - Ankle (R/L) - Great Toe (R/L)
Default setting is No.1.
Click
Save
to set the settings.
<Automatic Arterial Testing screen>