7.3 Instrument utilities
1.
Push
Home
to see the instrument model, version, serial number and location name.
2.
Push
Diagnostics
.
3.
Select an option.
Option
Description
Factory Service
For factory/service use only.
Instrument
Backup
Store
—Saves a backup of all the instrument settings and log files to a
USB flash drive.
Restore
—Copies the instrument settings and log files
from a USB flash drive to the instrument. Overwrites all the instrument
settings.
Instrument
Update
Installs an instrument update on the instrument from a USB flash drive.
Service Time
Shows the date entered for the last service date and for the next service
date. When set to on, a service reminder shows on the display when
service is due.
Section 8 Troubleshooting
Message
Solution
Startup
The self-check stopped.
Close the lid.
Close the lid. Push
Close
.
The self-check stopped.
Hardware error.
Set the power to off, wait 20 seconds and then set the power to on
again. If the self check is not successful, record the error number and
contact technical support.
Error numbers: 0: RTC; 1: Touch IC; 2: Slide door; 3: Dark voltage—
Close the door until a click is heard. Start the instrument again. 4:
Amplifier coefficient—Make sure that the power supply is connected to
an electrical outlet that has a protective earth ground. 5: Infrared LED
voltage; 6: Infrared LED current; 8: Transmission voltage drift—If the
lamp was replaced, calibrate the instrument. If a vial was in the sample
compartment during the self-test at startup, remove the vial. 9:
SDRAM; 10: NOR flash; 11: SPI flash; 12: Battery voltage; 13: Power
supply voltage—Make sure that the correct power supply is used.
Next calibration is due!
Calibrate the instrument. Refer to
Calibrate the turbidimeter with
on page 13.
Note: The calibration reminder is set to on. Refer to
Configure the calibration settings
on page 14.
Next service is due!
Contact technical support.
Note: The service reminder is set to on. Refer to
on page 21.
Next verification is due!
Do a calibration verification. Refer to the expanded user manual on the
manufacturer's website.
Note: The verification reminder is set to on.
Reading
Hardware error /
instrument error
Set the power to off, wait 20 seconds and then set the power to on
again. If the problem continues, contact technical support.