Option
Description
Live measure
This option is not available for all sensors. The option shows continuous
measurement data with connected sensors to quickly check
measurements. The installed applications and the automatic additions to
the sample are set to off. Continuous measurements are not
temperature compensated, so measurement differences may occur in
the same sample between continuous measurements and
measurements that use installed applications with temperature
compensation.
Reagent
replacement
Use this option to replace the reagents. Obey the instructions on the
display.
Maintenance
schedule
See the list of maintenance tasks. Push
Edit
to change the default
value.
Maintenance
summary
See the number of days remaining for maintenance tasks. After doing a
task, push
Reset
to set the number of days remaining to the default
value.
8.8 Storage and transportation
8.8.1 Prepare the instrument for storage
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
1.
Remove the sensor from the sensor holder and disconnect it from the sensor port on
the rear of the instrument.
2.
3.
Put the sensor in a storage tube that contains a storage solution. Refer to the sensor
documentation for storage instructions.
4.
From the home screen, select
Maintenance > Syringe activation
and empty the
contents of the syringe.
5.
If a second syringe is installed, push
Toggle
to select it and empty the contents.
6.
Set the instrument power to off.
7.
Carefully remove the tubes from the titrant bottle. If pumps are installed, remove the
tubes from the reagent bottles. Seal the bottles with the original bottle cap.
8.
Flush the tubes with water or an applicable cleaning solution (refer to the application
note for more information).
9.
Drain the used beakers and clean with deionized water.
10.
If storing the instrument for more than a week, remove the syringe from the
instrument (refer to
on page 12). Flush the syringe with water or an
applicable cleaning solution and store in its box.
Maintenance
46
Summary of Contents for TitraLab AT1102
Page 1: ...DOC022 52 93074 TitraLab AT1000 series workstations User Manual 03 2019 Edition 7...
Page 2: ......
Page 6: ...Specifications 4...
Page 20: ...Installation 18...
Page 25: ...Install accessories 23...
Page 26: ...Install accessories 24...
Page 34: ...Startup 32...
Page 42: ...Standard operations 40...
Page 50: ...Maintenance 48...
Page 58: ...Troubleshooting 56...
Page 64: ...Replacement parts and accessories 62...
Page 74: ...Change application parameters 72...
Page 79: ......