Appendix A Change application parameters
N O T I C E
The applications installed on the instrument have been pre-defined to optimize the measurement
process. Changing these default application parameters will have an effect on the measurement
process and measurement results. Only qualified personnel should change these parameters or
use the instrument in
Expert mode
. Refer to the application note before changing any
parameters.
N O T I C E
In
Basic mode
changes to parameters are secure. In
Expert mode
changes to parameters are
not secure and can cause the application to run incorrectly.
N O T I C E
The
Active
parameter is available to change in some parameter groups. It specifies if the method
should be used during the analysis. Changing this from
Yes
to
No
means that the method will not
be used which will have an effect on the measurement process and measurement results.
N O T I C E
The
Hide result
parameter (e.g. R2 hide, R3 hide, etc.) is available to change in some parameter
groups. It specifies if the result is shown on the display. Changing this from
Yes
to
No
means that
the calculated measurement result will not be shown. This will not have an effect on the
measurement process and measurement results.
The parameters available for change can be different for each instrument configuration
and the application being edited. The parameters marked with an "x" are available for
change.
Table 18 Application
Parameter
Description
Basic mode
Expert mode
Name
Name of the application
x
x
Advisable syringe
The application has been designed with this type of syringe
Table 19 Sample
Parameter
Description
Basic mode
Expert mode
Name
Name of the sample
x
x
Amount
Amount of the sample
x
x
Unit
Unit of the sample
x
Min. amount
Minimum permitted sample amount
x
Max. amount
Maximum permitted sample amount
x
Resolution
Number of decimals to show for the sample amount
x
Table 20 QC
Parameter
Description
Basic mode
Expert mode
Name
Name of the QC
x
x
Table 21 Blank
Parameter
Description
Basic mode
Expert mode
Description
Description of the blank
x
x
63
Summary of Contents for TitraLab AT1102
Page 1: ...DOC022 52 93074 TitraLab AT1000 series workstations User Manual 03 2019 Edition 7...
Page 2: ......
Page 6: ...Specifications 4...
Page 20: ...Installation 18...
Page 25: ...Install accessories 23...
Page 26: ...Install accessories 24...
Page 34: ...Startup 32...
Page 42: ...Standard operations 40...
Page 50: ...Maintenance 48...
Page 58: ...Troubleshooting 56...
Page 64: ...Replacement parts and accessories 62...
Page 74: ...Change application parameters 72...
Page 79: ......