9.4 Sterilization
W A R N I N G
Burn hazard. Obey safe handling protocols during contact with hot components. Wear
the appropriate protective equipment during maintenance or service activities.
The tip cone module can be sterilized by autoclaving it at 121 °C (250 °F) for 20 minutes. You can
use steam sterilization bags if needed. Disassemble and reassemble the tip cone module. Refer to
on page 10.
After autoclaving, the module must be cooled to room temperature for at least two hours. Before
pipetting, make sure that the module is dry. It is recommended to check the calibration after every
sterilization cycle.
Section 10 Troubleshooting
Problem
Possible cause
Solution
Leakage
Tip incorrectly attached
Attach firmly
Foreign particles between tip and tip
cone
Clean tip cones. Attach new tips.
Foreign particles between the piston,
the O-ring and the cylinder
Clean and grease the O-ring and
cylinder.
Insufficient amount of grease on
cylinder and O-ring
Grease accordingly
O-ring damaged
Change the O-ring.
Inaccurate
dispensing
Incorrect operation
Follow the instructions carefully.
Tip incorrectly attached
Attach firmly
Calibration altered: caused by
misuse, for example
Recalibrate according to the
instructions.
Inaccurate
dispensing with
certain liquids
Unsuitable calibration. High viscosity
liquids may require recalibration
Recalibrate with the liquids in
question.
Not dispensing
Pistons stuck or not connected
Remove the tip cone module. Move
the piston by hand or with piston
removal tool. Attach the module in
service mode.
English
11
Summary of Contents for TenSette plus
Page 101: ...4 1 4 1 1 4 1 2 101...
Page 103: ...5 5 1 1 1 2 1 3 1 4 1 5 2 TenSette plus Hach 103...
Page 104: ...LCD 1 2 5 3 60 2 2 1 2 5 4 3 106 5 5 1 2 3 4 104...
Page 106: ...3 1 4 7 2 5 8 3 6 9 7 MSDS SDS 7 1 1 Pipette 2 3 OK Speed 1 9 4 OK 5 OK 6 7 8 9 10 106...
Page 108: ...10 4 1 2 5 1 4 7 2 5 8 3 O 6 9 9 3 TenSette plus Hach Hach BBP088 9 4 121 C 250 F 20 107 108...
Page 109: ...10 O 109...
Page 170: ......
Page 171: ......