Figure 4 Relay function
1
Duration
3
Interval
2
Off delay
4
Time (x-axis)
Set the flow rate
1.
Push
MENU
.
2.
Select Test/Maint>Test Relay>Energize to do a flush.
3.
Slowly open the flow control valve to set the flow rate for the auto flush. Refer to
Make sure that the flow rate is sufficient to remove sediment but does not cause
water to come out of the turbidimeter.
4.
When the flow rate is correct, select De-energize to stop the flush.
Replacement parts
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate
distributor or refer to the company website for contact information.
Description
Item no.
Solenoid valve, 3-way, 120 VAC
2628400
Solenoid valve, 3-way, 230 VAC
2628402
Flow control valve, ½-inch NPT PVC
4668800
Reducer bushing, ¾-inch MPT x ½-inch FPT
4698400