Plumbing
Make the plumbing connections that are shown in
. The pipe, adapters and tee
fittings are supplied by the user.
Notes:
•
The valve location and orientation is not important.
installation.
•
The flow control valve fittings are ½-inch NPT.
•
The solenoid valve fittings are ¾-inch NPT. Install the supplied reducer bushings on
the solenoid valve fittings to connect ½-inch NPT pipe.
•
If the sample temperature is more than 50° C (122° F), use CPVC or other heat
resistant material (e.g., PVC pipe or flexible polypropylene) for all plumbing
connections made on the sample line to the solenoid valve.
•
Make sure to use pipe materials with a resistance rating that is sufficient for the
sample.
6