Page 60
Final Check List
8837prgwrk.fm
Section 6
6.5 Final Check List
For Storm Water Setup
•
Intake tubing must slope downward from the sampler to the source—no
loops or dips.
•
Does the pump create a lift? If not:
•
Did you secure the anchor strainer to prevent the flow velocity from
lifting the strainer to the water surface and exposing it to air?
•
Make sure the anchor strainer is submerged in channel.
•
Check all intake and pump tubing connections, especially the Teflon
Connection Kit (Cat. No. 2186) for vacuum leaks at the connection
points.
•
Did you calibrate the sample volume?
•
When the starting conditions occur, will the anchor strainer be
submerged? (i.e. when starting on a level of 0.5 inches).
If this example was for a circular pipe, the anchor strainer would not be
submerged because the Teflon/Stainless Steel anchor strainer is
0.84 inches in diameter.
Note:
A shallow depth, 0.406 in. diameter anchor strainer (Cat. No. 2071) is
available for low flow applications.
•
Does the distributor deposit samples in the bottle?
•
Press the
Start Program
key; this will cause the distributor to seek
bottle #1.
•
Press the
Change/Halt
key.
•
With the program halted, press the
Take Sample
key.
•
Press the
Bottle Advance
key.
•
Press the
Take Sample
key again.
•
Remove the controller from the bottle tray. Are samples in bottle #1
and #2?
•
If the sampler is being powered by a battery, is the battery fully charged?
•
Are the knobs on the liquid sensor and pump cover hand tight?
•
Make sure the sampler is level.
•
Before leaving, make sure the display reads
PROGRAM RUNNING
.
Summary of Contents for SIGMA 900
Page 2: ...Visit http www hach com...
Page 14: ...Visit http www hach com...
Page 22: ...Visit http www hach com...
Page 26: ...Visit http www hach com...
Page 32: ...Visit http www hach com...
Page 34: ...Visit http www hach com...
Page 64: ...Visit http www hach com...
Page 84: ...Visit http www hach com...
Page 86: ...Visit http www hach com...
Page 88: ...Visit http www hach com...
Page 94: ......