Calibration by comparison/grab sample

The sensor can be calibrated by measuring the process water with a portable instrument or 
taking a grab sample and measuring the value in the lab (one-point calibration). The sensor 
does not need to be removed from the process during calibration. The sample temperature 
should correspond to the measured process temperature for accurate results.

Note: 

The calibration by comparison option is not available when the range and units are set 

to 000.0%.

1. 

Press 

CAL

, enter passcode: 

1105

ENTER

. The display will show CAL PRD and then StorE. 

2. 

Measure the process water with a portable instrument or collect a grab sample for mea-
surement in the lab. Press 

ENTER

 to proceed. 

The cal mode indicator will flash to indicate that calibration has not been completed. The 
transmitter will continue to measure and display the current sample values.

3. 

Measure the grab sample with a laboratory or portable meter.

4. 

Press 

CAL

, enter passcode: 

1105

ENTER

 to access the product calibration once more. 

The display will briefly show CAL PRD and then CALC.

5. 

Edit the displayed value to match the value measured by the laboratory or portable meter 
and press 

ENTER

. The display will show the new cell constant. 

6. 

Press 

ENTER

 to end the calibration. The display will alternate between the measured value 

and Hold. 
Press 

ENTER

 to return to the measuring mode. The outputs will remain in the hold mode for 

approximately 20 seconds.

Note: 

For comparison calibration via Foundation Fieldbus, see the User Manual.

Table 1 Default passcode settings

Display

Function

Default setting

Cal INFO

Calibration information

0000

CAL 0-CAL

Calibration zero

1001

CAL CELL

Calibration—cell factor adjustment

1100

CAL SOL

Calibration with standard solution

0110

1

CAL PROD

Product calibration

1105

1

CAL RTD

Temperature sensor adjustment

1015

CFG ERR

Error information

0000

CFG CONF

Configuration mode

1200

CFG SNSR MO

Sensor monitor

2222

CFG OUT

Current source

5555

CFG SPCL ST

Administrator passcode. Use 

ARROW

 keys to 

select NO or YES

1989

NO SPCL ST

Do not change 
administrator 
passcode

Press 

ENTER

 to return to the default 

Press 

CONF

 to exit the editor

YES SPCL ST

Change administrator 
passcode

Press 

ENTER

 to accept the new passcode

Press 

CONF

 to exit the editor

1

Not available with si792x E/T-PA