Option Description
rCL
Shows the last 10 measurements recorded. Push to show the
recorded measurements (01—most recent measurement, 10—
oldest measurement). Push to scroll through the measurements.
To select a measurement by number, push to select the number
and then . Push to go out of this option.
SCA
Refer to the expanded user manual on the manufacturer's website.
3.
Push to go back to measurement mode.
Run a test
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and
wear all of the personal protective equipment appropriate to the
chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
The basic measurement steps necessary to run a test follow. Refer to
the applicable method to run a specific test.
1.
Select the applicable measurement range or parameter. Refer to
on page 11.
2.
Prepare the blank according to the method document. Make sure to
use the correct sample cell size. Rinse the sample cell and cap with
the blank before the sample cell is filled.
3.
Close the sample cell and clean the optical faces of the sample cell
with a lint-free cloth.
4.
Insert the blank sample cell into the cell holder. Make sure to install
the blank sample cell in the correct and consistent orientation so
that the results are more repeatable and precise. Refer to
.
5.
Install the instrument cap over the cell holder. Refer to
6.
Push to set the instrument zero. The display shows "0.000", or
the degree of resolution that was previously selected.
7.
Prepare the sample. Rinse the sample cell and cap with the sample
three times before the sample cell is filled. Add reagents as
specified by the method document.
12
English
Summary of Contents for POCKET COLORIMETER II
Page 2: ...English 3 e tina 20 37 Magyar 56 Rom n 73 Slovensk jazyk 90 Slovenski 107 Hrvatski 124 2...
Page 8: ...Figure 2 Install the batteries 8 English...
Page 25: ...Obr 9 Instalace bateri e tina 25...
Page 39: ...39...
Page 41: ...15 1 2 3 15 41...
Page 42: ...15 1 3 1 cm1 5 2 4 1 16 42...
Page 43: ...16 43...
Page 44: ...17 17 18 44...
Page 45: ...18 1 1 3 2 4 19 45...
Page 46: ...19 1 4 2 5 3 6 10 1 2 SEL 46...
Page 47: ...00 00 24 rCL 10 01 10 SCA 3 MSDS SDS 1 46 2 3 4 47...
Page 48: ...20 5 21 6 0 000 7 8 9 20 10 21 11 12 13 48...
Page 49: ...20 1 2 25 mm 10 mL 3 1 cm 10 mL 49...
Page 50: ...21 rCL 46 50...
Page 51: ...MSDS SDS Liquinox 51...
Page 52: ...42 E 0 E 1 1 E 2 LED1 LED LED 52...
Page 53: ...E 3 E 6 Abs E 7 E 9 53...
Page 54: ...2 1 E 1 E 2 _ __ E 1 E 2 2 10 54...
Page 55: ...4 pkg 4674300 1 5955900 1 5954800 25 mm 10 mL 6 pkg 2427606 1 cm 10 mL 2 pkg 4864302 55...
Page 84: ...Figura 34 Orientarea cuvei 1 Marcaj de orientare 2 Cuv 25 mm 10 ml 3 Cuv 1 cm 10 ml 84 Rom n...
Page 112: ...Slika 44 Namestitev baterij 112 Slovenski...
Page 129: ...Slika 51 Umetanje baterija Hrvatski 129...
Page 141: ......