6
English
Operation
Settings (continued)
Measuring
Interval
Enter how often measurements are to be taken. ATTENTION with filtration probe and 5min
operation: increased pump speed in the filtration probe, annual filtration probe maintenance is
necessary instead of every 2 years.
Set to
WET/DRY, decides if the cuvette is empty (DRY) or filled (WET) between measurements at
measurement intervals that are equal or greater than 10 minutes. Adapt individually for
smoothest measurement values.
Start by bus
Start by bus
YES/NO, decides if the instrument is measuring continuously or measurements are triggered
by field-bus. The "Fieldbus" option must be activated in the Test/Maintenance menu. The
instrument will switch to a 5 min interval when activated.
Measurements
Number of measurements that are taken after one activation by bus.
Discarded
measurements
Number of discharged values that precedes the measurements.
Average
Number of measurements that are averaged. (effects bus triggered measurements only)
Cleaning
Interval
Number of hours between the cleanings.
Start
Start time for the cleaning (in case of more than one cleaning per day: start time for the first
cleaning)
Discarded
measurements
Number of measured values that are discarded after a cleaning operation.
Output mode
Value output during a cleaning operation and the following discarded values. HOLD = last
measured value, SET TRANSFER = value to be entered
Tube heating
On
Probe tube heating switches on at the start of the selected month. With filtration probe.
Off
Probe tube heating switches off at the end of the selected month. With filtration probe.
Reagent warning
Reagent warning
On/Off: On determines the warning output if the reagent levels are low.
Warning
Determines the level below which the reagent must drop to trigger the warning.
Module status: Warning
40 %, 30 %, 15 %
Warning is released, when a filter probe is installed and the status of filtration modules falls
below the defined level.
Module status: Error
14 %,10 %,8 %, Off
Status of filtration modules at filter probe at which an error is generated. When switched to
OFF, a deactivated sample detection will be switched to "warning".
Sample detection
Off/Warning/Error
Determines instruments reaction when amount of available sample is to low. When
instrument is in filter probe mode, deactivating the sample detection will switch a deactivated
"STATUS MODUL.ERR" to 14 %
Drain blockage
On/Off
Determines instrument reaction when drain is blocked
Reference electrode
On/Off
Determines if measurement reference channel is used for measurement or not.
Bubble rejection status
On/Off
Use for samples that produce gas when acid is added. If active, sample is removed from
cuvette after reagent has been added, then refilled again to remove bubbles. If active, the
measurement interval of 5 minutes is not available. For not degassing samples the
measurement values are smother with bubble rejection deactivated.
5.2 Sensor setup menu (fortgesetzt)
Summary of Contents for PHOSPHAX indoor sc
Page 2: ......
Page 4: ...4 Table of contents...
Page 20: ...20 English Troubleshooting...
Page 38: ...38 Deutsch Fehlersuche und Behebung...
Page 88: ...88 Italiano Individuazione ed eliminazione dei guasti...
Page 104: ...104 Nederlands Problemen oplossen...
Page 120: ...120 Dansk Fejlfinding...
Page 138: ...138 Polski Rozwi zywanie problem w...
Page 139: ......