65
Inputs/Outputs Menu
8.5 Analog output characteristics
Analog output characteristics
“Linear” analog output
The "Linear" output is the default setting for the analog output. It is
illustrated below (4-20 mA output is shown, 0-20 mA settings are similar).
The goal of this setting is to use all the points available on the slope from
4 mA to 20 mA to show the range of measurements that are usual in the
measured process. Setting the output this way allows for the highest signal
resolution for the actual conditions.
The downside is that any measures below the set range will have the same
analog signal locked at 4 mA. Similarly, any measure over the set range will
have the same analog signal locked at 20 mA. Settings must be made in
balancing these aspects.
Settings
For the output, set ML and MH in the current measuring unit (e.g. °C for a
temperature output). When a compound unit is selected the smallest unit will
be used (e.g. ppb for a "ppm-ppb" compound unit).
These points should be set keeping in balance the following conditions (see
illustration above):
•
The smaller is Range 1, the better is the analog signal resolution within
the expected range of measure.
•
In Range 0 the analog output only shows that measurement is below the
ML value. Similarly in Range 2 the analog output only shows that
measurement is over the MH value.
The formula to compute the measurement knowing the current I and the
resolution R is given in the following table:
Linear
Range
Measurement M
Resolution R
4-20 mA
20 > I > 4
M = ML + (MH - ML)
*
(I - 4) / 16
R = (MH-ML) / 808
0-20 mA
20 > I > 0
M = ML + (MH - ML)
*
I / 20
R = (MH-ML) / 1010
Summary of Contents for ORBISPHERE K1200
Page 5: ...4 Table of Contents...
Page 13: ...12 General Information...
Page 39: ...38 Installation...
Page 45: ...44 User Interface...
Page 46: ...45 Section 5 View Menu Figure 34 View menu...
Page 49: ...48 View Menu...
Page 59: ...58 Calibration Menu...
Page 60: ...59 Section 8 Inputs Outputs Menu Figure 40 Inputs Outputs menu...
Page 62: ...61 Inputs Outputs Menu 8 4 Analog outputs Figure 41 Analog outputs menu...
Page 69: ...68 Inputs Outputs Menu...
Page 86: ...85 Section 11 Products Menu Figure 44 Products menu...
Page 89: ...88 Global Configuration Menu...
Page 90: ...89 Section 13 Services Menu Figure 46 Services menu Part 1...
Page 91: ...90 Services Menu Figure 47 Services menu Part 2...