Observation
Explanation
The bottle goes up with the
piercing head
Stop the analysis by pressing the emergency STOP button.
Disconnect instrument from the power supply and remove the bottle. Check
that the piercing spike is in place and not unscrewed. Release emergency
STOP button by turning it clockwise and restart the instrument.
Open the front door to stop the process. Remove the bottle and check that
the piercing spike is in place and is not unscrewed. Close the front door and
press the
Abort
button on the screen.
No antifoam being injected
Go to the antifoam section of the maintenance wizard and prime the pumps
until antifoam can be seen being regularly injected into the antifoam priming
beaker.
26