Specifications
Specifications are subject to change without notice.
Analyzer
Specification
Details
Operating temperature
–5 to 50°C (23 to 122 °F)
Storage temperature
–20 to 70°C (–4 to 158 °F)
Operating humidity
0 to 95% non condensing relative humidity
Operating altitude
From 0 to 2,000 m. (6,550 ft.) above sea level
EMC requirements
EN61326-1: EMC Directive
Note: The wall mount instrument is a Class A product. In a domestic environment this product
may cause radio interference in which case the user may be required to take adequate
measures.
Korean registration
User Guidance for EMC Class A Equipment
업무용을
위한
EMC
등급
A
장치에
대한
사용자
지침
사용자안내문
A
급
기기
(
업무용
방송통신기자재
)
이
기기는
업무용
(A
급
)
전자파적합기기로서
판매자
또는
사용자는
이
점을
주의
하시기
바라며
,
가정외의
지역에서
사용하는
것을
목적으로
합니다
.
CE compliance
EN61010-1: LVD Directive
Safety rating
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
Enclosure ratings
IP 65; Totally protected against dust; Protected against low pressure jets of water
from all directions.
NEMA 4X (wall mount only); Totally protected against dust; Protected against
pressure jets of water from all directions.
W A R N I N G
Enclosure rating does not apply to external power supply for benchtop
instruments.
Power supply
Universal 100 VAC to 240 VAC @ 50/60 Hz - 40VA; 10 to 30 VDC - 30W
Analog current output version
on the measurement
board(s)
4-20 mA (default) or 0-20 mA (configuration with software);
3 configurable outputs: Maximum load: 500 ohm; Sensitivity: 20μA;
Accuracy: ± 0.5% (between operating temperature limits)
Analog voltage output
version on the measurement
board(s)
0- 5 V output (hardware option);
3 configurable outputs; Minimum load: 10 KOhm; Sensitivity: 5 mV;
Accuracy: ± 0.5% (between operating temperature limits)
Measurement alarm relays
on the measurement
board(s)
Three alarm relays per measurement board; 1A-30 VAC or 0.5A-50 VDC on a
resistance load.
Configurable to Normally Open [NO] or Normally Closed [NC] contacts by
changing the jumper positions.
W A R N I N G
Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC
RMS
English
3