Connections to electronic boards
N O T I C E
Potential Instrument Damage. Delicate internal electronic components can be damaged by static electricity,
resulting in degraded performance or eventual failure.
Note: Any loose connection wires should be bundled tightly together with the use of nylon cable ties.
Sensor cable
An ORBISPHERE cable (10 wire shielded, Part N° 32505.mm) is needed to connect the sensor(s) to
the instrument. The instruments have a Lemo 10 socket on the back panel where the sensor cable
has to be connected.
Electronic boards connectors
Connectors P8 on the main board (
on page 12) and J7 on the measurement board (refer
to
51x measurement board
) are made of two parts. Push down carefully the black levers on either
side of the connector and pull it out securely. Perform all connections with these connectors
unplugged. Once finished, attach the connectors to the boards by pushing them firmly in place
(levers up).
Main board
Figure 4 Main board
Figure 5 Connector P8
Connector P8
The numbers listed below refer to the 13 available P8 connections (from left to right) in
1. RS-485 (signal A)
2. RS-485 (signal B)
3. PROFIBUS-DP (GND)
4. PROFIBUS-DP (+ 5 V)
5. PROFIBUS-DP (signal -)
6. PROFIBUS-DP ()
7. PROFIBUS-DP (signal RTS)
8. Not used
9. Not used
10. Not used
11. System alarm relay (N.O.)
12. System alarm relay (N.C.)
13. System alarm relay (Common)
Connector P3
Ethernet RJ 45. Connect the instrument to the local network by passing an ethernet cable through
the ethernet cable gland (refer to
on page 10) and connecting to the
.
12