Services

A number of maintenance and diagnostic options
are available from the 

Services

 menu.

Sensor Diagnostics - Calibration Timer

The instrument can automatically remind the user
when the next sensor calibration is due. Select
enable and enter a delay in days. The display
shows current instrument date and time, next
calibration due date and time, and remaining
days. 
The next calibration date is updated when the
sensor is calibrated. 

Sensor Diagnostics - Service Timer

The instrument can automatically remind the user
when the next sensor service is due. Select
enable and enter a delay in days. This should be
validated by a level 3 user. The display shows
current instrument date and time, next calibration
due date and time, and remaining days.
The next service date is updated when the button

Service done

 is pressed after a service. 

The sensors attached to your instrument will
require periodic servicing and maintenance. For

more information on this, please refer to the
corresponding sensor maintenance manual.

Language

Check the language as required and restart the
instrument to apply the change. The instrument
will restart in the language selected. 

Other Options

These include:

Time/Date

 - Used to set the time, date 

and the display settings

Screen

 - Used to set the screen contrast

Buzzer

 - Used to set the alarm 

parameters (sound, volume, etc.)

Main board info

 - Displays instrument 

model, ID and software version

Batteries

 - This display gives the real 

time clock battery charge level and 
voltage

Instrument Maintenance

Any instrument maintenance should be carried
out by a qualified Hach Ultra Service Technician.
Please contact your local representative should
yo u f ee l an y ma in te na nc e o r in st ru m en t
adjustments are required.

Troubleshooting - List of Events

The possible events, along with the text message
displayed on the instrument numeric view screen,
the reason for the event and its criticality are listed
in the following table. 
An  event is something which affects the
measurement. 

Maintenance and Troubleshooting

Summary of Contents for ORBISPHERE 410

Page 1: ...Operator Quick Guide ORBISPHERE 410 Revision E 24 04 2008...

Page 2: ...interface software is Windows CE based providing easy selection through menus All the measurement configuration calibration and standard service routines can be called by pressing buttons and menu bar...

Page 3: ...f information displayed depends on the gas being measured and the channel configuration The statistical view offers statistical data that matches with Total Quality management tools Statistics is a to...

Page 4: ...lved Gas unit The list of available units depends on unit type selected above Liquid When medium is liquid select water or a liquid with a different solubility if available Display resolution Maximum...

Page 5: ...ntration values are above High alarms or below Low alarms Set this to a minimum but enough to avoid alarms for non representative peaks beyond the set level Measurement Filter Configuration The filter...

Page 6: ...the buffer is full Adjust the parameters for recording and storing measured data Storage mode selection No storage Store once When volatile memory is full 1 000 positions measurement recording stops R...

Page 7: ...the Modify button Last calibration parameters are memorized so this step can be ignored if correct parameters are already set Modify Calibration Parameter Calibration mode 2 types available depending...

Page 8: ...saturated air to provide a known oxygen reference against which to calibrate Dry the sensor thoroughly before placing the sensor storage cap under tap water Shake off any excess water but leave a few...

Page 9: ...tion temperature barometric pressure current the date and time the user name and ID Barometric Pressure Calibration Upper box shows the barometric pressure measured by the instrument Using a precision...

Page 10: ...cing and maintenance For more information on this please refer to the corresponding sensor maintenance manual Language Check the language as required and restart the instrument to apply the change The...

Page 11: ...urement is frozen Alarm low low The gas concentration is below the Alarm LowLow limit Alarm low The gas concentration is below the Alarm Low limit Alarm high The gas concentration is above the Alarm H...

Page 12: ......

Reviews: