Specifications
Specifications are subject to change without notice.
Instrument specifications
Specification
Detail
Power requirement
Adapter (included in the ship kit): 100–240 VAC, 2.5 A, 50–60 Hz
Instrument: 24 VDC, 75 W maximum
Installation category
I
Protection class
III
Pollution degree
2
Altitude
2000 m (6562 ft)
Light source
Long Life Laser
™
diode with 10-year Mean Time To Failure (MTTF), Class 3B Laser,
810 to 852 nm, 50 mW maximum
Pump type
Air vacuum, rated for continuous use
Count display
Color ¼ VGA TFT touch screen
Interface
Windows CE
®
-based
Maximum count shown 9,999,999
Delay time
00:00:06 to 23:59:59
Sample and hold times Sample: 00:00:01 to 23:59:59
Hold: 00:00:00 to 23:59:59
Count alarms
1 to 9,999,999 counts
Data storage
50 to 5000 samples, scrollable on Historical Data review screen
3000 is the default value
Count cycles
Up to 100 while in automatic mode
Locations
Up to 999
Exhaust port
3/8-in. NPT thread
Outputs
Ethernet–10Base-T/100Base-TX
RS485 Serial
RS232 Serial
Optional wireless–802.11 b/g compatible
USB Client (Version 1.1)
USB Host (Version 1.1)
Manifold
Supports A3432, 32-port manifold system (available on 1 CFM units only)
Enclosure material
Stainless steel
Weight without battery
3413 and 3415—7.55 kg (16.6 lb)
3423 and 3425—8.33 kg (18.3 lb)
3445—8.65 kg (19.0 lb)
Size (W x D x H)
31.8 x 25.4 x 20.3 cm (12.5 x 10 x 8 in.)