Run Test
7
RUN TEST
This section tells you how to run a test or QC test. The examples show screen shots for a
LightDeck MC + CYN Test Cartridge.
If cartridges were refrigerated, allow to warm to room temperature for at least 15 minutes.
Start instruments and software
1. Turn on the computer
2. Start LightDeck-T software
3. Double click the LightDeck-T icon on the desktop to launch the software.
Test Procedure
Click the RUN TEST button.
Enter sample information.
•
Choose a test from the drop-down
menu.
•
Enter Sample Information
•
Click ACCEPT to load your inputs.
Open the pouch. Remove the test
cartridge.
•
Avoid touching the clear plastic on the
bottom of the test cartridge.
•
Label the test cartridge with a sample
ID.
Caution:
A desiccant sachet should also
be in the pouch. Discard the cartridge if
the desiccant sachet is not present.