![Hach Lachat BD40HT User Manual Download Page 19](http://html1.mh-extra.com/html/hach/lachat-bd40ht/lachat-bd40ht_user-manual_3195231019.webp)
Table 5 Safety messages (continued)
Message
Possible cause
Solution
Temperature Dropping
The sample may have evaporated.
Make sure that there is sample in the
digestion tubes.
An endothermic reaction has
occurred.
Make sure that the method in the controller
matches the written method. Repeat the
test.
Check the Dongle
The dongle is not connected.
Put the dongle in the COM port.
Set Point Exceeds Sample
Boiling Point
The temperature set point is higher
than the sample boiling point.
Change the temperature set point.
The rate of evaporation of the sample
is too high.
Change the temperature set point.
Temperature Unsteady
The sample has evaporated.
Make sure that there is sample in the
digestion tubes.
Replacement parts and accessories
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Quantity
Item no.
Heat deflector shield
2/package
18091L
Power cord, North America, 230 V, 15 A with C19 connector
each
18092L
Test tube and cooling rack, 40-position
each
18093L
Cold finger set
40/package
18032L
Digestion tube set, 25 mL and 50 mL, round bottom
40/package
18094L
Replacement dongle for controller
each
18095L
Fuse, 15 A, 3 AB
2/package
18105L
Accessories
Description
Item no.
Calibrated test tube, 25 mL and 50 mL, round bottom
18090L
Cold finger, for calibrated digestion tubes
18032
Boiling chips, 250 g
18501
English
19
Summary of Contents for Lachat BD40HT
Page 2: ...English 3 Fran ais 20 Espa ol 38 Portugu s 56 73 89 106 2...
Page 35: ...Fran ais 35...
Page 53: ...Espa ol 53...
Page 74: ...74...
Page 76: ...1 1 5 2 6 3 COM 2 7 4 2 1 SiC 5 40x 2 6 3 2x 7 4 40x 76...
Page 77: ...3 3 77...
Page 78: ...4 4 GFCI GFI 5 5 1 230 V 15 A C19 78...
Page 79: ...6 1 6 1 5 2 6 3 7 4 8 1 7 2 79...
Page 80: ...7 1 2 3 2 Teflon 260 C DigiTUBE HT 1 5 78 2 1 76 3 80 81 MSDS 80...
Page 81: ...1 2 3 4 COM 2 1 76 5 6 7 8 8 8 1 2 3 4 1 5 81...
Page 85: ...1 2 3 4 1 2 3 4 5 6 4 5 7 8 85...
Page 86: ...86...
Page 88: ...25 mL 50 mL 18090L 18032 250 g 18501 88...
Page 90: ...90...
Page 92: ...BD40HT 1 12 40 2 450 C 1 1 5 2 6 3 COM 2 7 4 92...
Page 93: ...2 1 SiC 5 40 2 6 3 2 7 4 40 3 93...
Page 94: ...3 2 4 94...
Page 95: ...4 GFCI GFI 5 5 1 230 V 15 A C19 95...
Page 96: ...6 1 6 1 5 2 6 3 7 4 8 1 7 2 96...
Page 97: ...7 1 2 3 2 260 C DigiTUBE HT 1 5 95 2 1 92 3 98 98 MSDS 97...
Page 98: ...1 2 3 4 COM2 1 92 5 6 7 8 8 8 1 NEW METHOD 98...
Page 102: ...1 2 3 3 2 1 2 3 4 1 102...
Page 103: ...2 3 4 5 6 4 5 2 3 7 8 103...
Page 104: ...5 CONTINUE SUSPEND 104...
Page 107: ...107...
Page 109: ...1 1 5 2 6 3 COM 2 7 4 2 1 SiC 5 40x 2 6 3 2x 7 4 40x 109...
Page 110: ...3 3 110...
Page 111: ...2 4 4 GFCI GFI 5 111...
Page 112: ...5 1 230V 15A C19 6 1 6 1 5 2 6 3 7 4 8 112...
Page 113: ...1 New Method 7 2 7 1 2 3 2 260 C DigiTUBE 1 HT 1 5 112 2 1 109 113...
Page 114: ...3 114 New Method 115 New Method MSDS 1 2 3 4 COM 2 1 109 5 6 7 8 8 114...
Page 118: ...6 7 8 TempPROBE 9 4 7 10 1 2 3 1 2 3 4 118...
Page 119: ...1 2 3 4 5 6 4 5 7 8 119...
Page 120: ...120...
Page 122: ...1 18095L 15A 3AB 2 18105L 25mL 50mL 18090L 18032 250 g 18501 122...
Page 123: ......