Hach ISENH4181 User Manual Download Page 13

Table 2   Troubleshooting information (continued)

Problem

Possible cause

Solution

Sample properties cause slow
stabilization or inaccurate
measurements.

The sample pH with ISA is
incorrect.

Make sure that the sample pH
after the ISA is added is less
than pH 8.5. Make sure to add
one ISA pillow to each 25 mL of
sample.

The sample temperature is
low, or there is a large
temperature difference
between samples.

Increase the sample temperature
or adjust the temperature of
different samples to be the same
(within 2 °C (3.6 °F)).

Procedure problem causes slow
stabilization and prevents
accurate calibrations or
measurements.

Air bubbles are around or
below the probe tip.

Carefully tap or shake the probe
to remove air bubbles.

The ISA was not added.

Add one ISA powder pillow to
each 25 mL of sample and
standard solution.

The stir speed is too slow or
too fast.

Try a different stir speed.

Magnetic stirrers can
become warm and increase
the solution temperature.

Put a piece of insulating material
between the stirrer and beaker.

An incorrect standard
solution was used or the
standard solution has
contamination.

Use the specified standard
solution of good quality.

Too much time passed after
the ISA was added.

Measure the sample or standard
solutions immediately after the
ISA is added.

9.1 Slope check

Use the mV value of two standard solutions to make sure the probe gives the correct slope.

1.

Prepare two standard solutions that are ten times apart in concentration (e.g., 10 mg/L and
100 mg/L NH

3

—N). Select standard solutions with a concentration above and below the typical

sample concentration. Use a minimum concentration of 1.8 mg/L.

2.

Use the measurement procedure to add the ISA and measure the mV value of each standard
solution.

3.

Calculate the difference in the mV value of the two standard solutions to find the slope. If the
probe is in good condition, the slope will be 58 mV (within the ± slope limits of the method) at
25 °C (77 °F).

9.2 Standard additions check

To make sure that the sample measurement is accurate, add a small volume of a standard solution
to the sample and calculate the percent recovery. The sample with the known volume of standard
solution is known as a spiked sample.

1.

Use the measurement procedure to measure the concentration of a 25-mL sample.

2.

Use a pipet to add the applicable volume of standard solution to the sample. Refer to 

Table 3

.

English

   

11

Summary of Contents for ISENH4181

Page 1: ...DOC022 53 80029 ISENH4181 05 2021 Edition 5 User Manual...

Page 2: ......

Page 3: ...n 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 6 3 Low level measurements 8 6 4 Interferences...

Page 4: ...Table of Contents 2...

Page 5: ...2 Specifications Specifications are subject to change without notice Specifications Details Probe type Digital combination probe with a non refillable reference junction and a built in temperature sen...

Page 6: ...r Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory The Intellic...

Page 7: ...probe before use To condition the probe soak the probe in approximately 25 mL of the lowest concentration standard solution used for calibration for 30 to 60 minutes Do not add the ISA to the standard...

Page 8: ...e 1 Go to the calibrate menu Select the probe if applicable The display shows the standard solutions to use for calibration 2 Add 25 mL of each standard solution to different beakers 3 Add one Ammoniu...

Page 9: ...lete traceability is necessary enter a sample ID and operator ID before measurement Refer to the meter manual for instructions Stir the standards and samples at a slow and constant rate to prevent the...

Page 10: ...ow apply The sample does not contain interferences The sample pH is within the range given in the specifications Omission of the ISA is accepted by the regulatory reporting agency if the measurement i...

Page 11: ...o the acceptance criteria 7 1 Verification procedure 1 Go to the verification menu The display shows the standard solution to use for verification Note Menu name for HQd meters Run check standard 2 Po...

Page 12: ...rd solution 8 2 Storage Keep the sensor protection cap on the probe during storage to prevent damage to the sensing element Note Make sure to condition the probe after long term storage Refer to Prepa...

Page 13: ...contamination Use the specified standard solution of good quality Too much time passed after the ISA was added Measure the sample or standard solutions immediately after the ISA is added 9 1 Slope ch...

Page 14: ...tandard addition A percent recovery of 100 5 is an indication that the sample measurements are accurate Percent recovery CM CE x 100 Where CM measured concentration of the sample after the addition of...