Hach Intellical PHC805 User Manual Download Page 11

6.2 Sample measurement notes

Read the notes that follow before sample measurements.

• Rinse the probe with deionized water and dry with a lint-free cloth between measurements to

prevent contamination.

• If complete traceability is necessary, enter a sample ID and operator ID before measurement.

Refer to the meter manual for instructions.

• The meter automatically saves the measurement data when the user manually reads each data

point and when the meter is set to read at regular intervals. The user must manually save each
data point when the meter is set to read continuously.

• Air bubbles below the sensor can cause a slow response or error in the measurement. Make sure

to remove air bubbles before and during measurements.

6.3 Sample measurement procedure

1. 

 Collect the

sample.

2. 

 Open the filling

hole.

3. 

 Rinse the probe

with deionized water.
Dry the probe with a
lint-free cloth.

4. 

 Put the probe in

the sample with the
sensor and
reference junction
fully in the sample.
Do not put the probe
on the bottom or
sides of the beaker.

5. 

 Shake the probe

from side to side to
refresh the reference
junction and remove
air bubbles.

Note: Be careful to not
hit the beaker and
damage the probe.

6. 

 Stir gently, then

read the pH value of
the sample. The
display shows the
temperature-
compensated pH
value when the
reading is stable.

Section 7 Verify the calibration

Measure the pH value of a fresh pH buffer solution to make sure the result is accurate. The meter
compares the selected pH buffer value to the measured pH value and accepts or rejects the
measurement. The user can change the pH buffer solution and acceptance criteria for verification in
the probe-specific settings.

Note: Password protection may prevent access to the acceptance criteria.

English

   

9

Summary of Contents for Intellical PHC805

Page 1: ...DOC022 53 80436 PHC805 05 2021 Edition 2 User Manual...

Page 2: ......

Page 3: ...7 5 2 Calibration procedure 7 Section 6 Sample measurement 8 6 1 Sample requirements 8 6 2 Sample measurement notes 9 6 3 Sample measurement procedure 9 Section 7 Verify the calibration 9 7 1 Verifica...

Page 4: ...Table of Contents 2...

Page 5: ...mples of low ionic strength For best accuracy do not use the probe to measure the pH of very alkaline samples Figure 1 Probe overview 1 Probe soaker bottle with storage solution 5 Probe soaker bottle...

Page 6: ...acturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quali...

Page 7: ...ion for use N O T I C E Make sure to remove the protective tape from the filling hole of new probes A probe with a blocked filling hole will not operate correctly New probes are filled with electrolyt...

Page 8: ...Figure 2 Remove the protective tape Figure 3 Probe position during use 1 Open filling hole 3 Level of calibration solution or sample 2 Level of electrolyte filling solution 6 English...

Page 9: ...more than one probe is connected to the meter if applicable Calibrate the probes and verify the calibration regularly for best results Use the meter to set calibration reminders The calibration data...

Page 10: ...e meter documentation for meter operation and probe specific settings 6 1 Sample requirements N O T I C E Make sure that the sample type is compatible with the probe or probe damage can occur Samples...

Page 11: ...mple 2 Open the filling hole 3 Rinse the probe with deionized water Dry the probe with a lint free cloth 4 Put the probe in the sample with the sensor and reference junction fully in the sample Do not...

Page 12: ...ionized water Dry the probe with a lint free cloth 5 Put the probe in the pH buffer solution with the sensor and reference junction fully in the solution Do not put the probe on the bottom or sides of...

Page 13: ...ater Dry the probe body with a lint free cloth 4 Soak the probe in pH 4 buffer for 20 minutes 5 Rinse the probe with deionized water Dry the probe body with a lint free cloth Table 1 Cleaning solution...

Page 14: ...eplace the filling solution 1 Tilt the probe and open the filling hole 2 Use a plastic transfer pipet to remove the contaminated solution from the filling hole Discard the solution 3 Rinse the inner p...

Page 15: ...s of low ionic strength For short term storage the probe can stay in the sample for a maximum of 2 hours if the sample pH is not high Close the filling hole and put the soaker bottle that contains the...

Page 16: ...ilization or inaccurate measurements The sample absorbs carbon dioxide CO2 from the air which causes the pH value to slowly decrease in low ionic strength LIS or high purity samples Use the LIS chambe...

Page 17: ...ckets 20 mL each 20 pkg 2770020 pH 7 00 buffer solution Singlet one use packets 20 mL each 20 pkg 2770120 pH 10 01 buffer solution Singlet one use packets 20 mL each 20 pkg 2770220 pH 4 01 and pH 7 00...

Page 18: ...escription Quantity Item no Beaker 30 mL plastic colorless 80 pkg SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker...

Page 19: ......

Page 20: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Reviews: