5.
Click
ADD NEW
. The Add Instrument window opens.
6.
Enter the serial number (SN) of the logger (
7.
Select the Instrument Type.
8.
Select the Active check box and click
OK
. The instrument is shown in the Instrument Manager.
Configure the logger for remote communication
Pre-requisites:
The logger must be attached to the computer. An account with a network provider
must be set up, and the server must be configured.
The settings for remote communication must be entered into FSDATA Desktop and then written to
the logger.
1.
Start a communication session with the logger:
a.
Open FSDATA Desktop.
b.
Click
CONNECT
. The Connect to Instrument window opens.
c.
Click the FL900 button.
d.
Select the port on the computer where the logger is attached (serial or USB), then click
Connect
.
Note: If the sensor mismatch message is shown, select "Create new program based on sensors
connected."
e.
Make sure that the connection status shows "connected".
2.
Go to the Communications tab. Enter the information for the desired site. Select Time Zone.
3.
Complete the Remote Settings information:
Option
Description
Verizon 4G
No additional configuration is necessary.
LAT3 4G
Select the network provider and enter the user name and password, if applicable.
Primary Call Interval
The frequency that the logger calls the server.
Secondary call interval
The frequency that the logger calls the server during an alarm condition.
Server Verification Code
The code that specifies which account on the FSDATA server the logger belongs
to. The server cannot store data without this code.
4.
Click
WRITE TO LOGGER
to save the settings. A message window is shown:
Option
Description
Warning: all data
will be lost.
Continue?
All data that is stored in the logger is erased when a program is written to the logger.
To save the data, select No and download the data to a safe location. Select Yes to
erase all data and update the logger with the new program.
Set Logger Clock
Synchronize to Computer Clock—the logger uses the date and time settings of the
computer. Set Logger Clock—the logger uses the date and time settings that are set
by the user. If the unit has a modem, the logger automatically uses the date time
settings of the server.
Verify the telemetry (wireless option)
The user can manually send a call to the server to make sure that the network communication is
good.
1.
Temporarily attach the antenna to the logger to test the antenna and the cell coverage at the site
location before installation.
2.
Touch the magnet to the call initiation target (
). The modem LED indicator changes to
green.
3.
Look at the modem LED indicator during the call (45 to 90 seconds) and wait for a change:
• LED goes off—the call to the server was successful.
English
19
Summary of Contents for FL900 Series
Page 10: ...Figure 3 Battery cover removal 1 2 10 English...
Page 17: ...1 2 3 4 English 17...
Page 35: ...1 2 3 Deutsch 35...
Page 42: ...1 2 3 4 42 Deutsch...
Page 62: ...1 2 3 62 Italiano...
Page 87: ...1 2 3 Fran ais 87...
Page 94: ...1 2 3 4 94 Fran ais...
Page 113: ...1 2 3 Espa ol 113...
Page 139: ...1 2 3 Portugu s 139...
Page 163: ...Figur 3 Borttagning av batterilock 1 2 Svenska 163...
Page 188: ...ekil 3 Pil kapa n n kar lmas 1 2 188 T rk e...
Page 195: ...1 2 3 4 T rk e 195...
Page 213: ...1 2 3 Slovenski 213...
Page 238: ...1 2 3 238 Hrvatski...
Page 255: ......