4.-6. TIC Span Adjust.

 This menu item allows the user to set the TIC span adjust factors manually for each 

range as described for TOC Span Adjust above. 

 

10.  Run Span Calibration. 

This function starts the Span Calibration cycle. The span calibration reactions are 

run at a single range programmed by the RANGE in this menu below. At the end of the span calibration 

cycle, BioTector automatically calculates the Span Adjust factors and displays it for Span Adjusts above. 

Unless it is manually modified, the same Span Adjust factor calculated for the programmed RANGE in 

this menu is used for the other two ranges as well. The span reactions operates in the same manner as a 

normal reaction, but the Sample Pump reverse operation is disabled to prevent the contamination of the 

standard solution connected to calibration/manual port. Span Calibration reactions have the prefix of 

“SC”.  

 

11.  Run Span Check. 

This function starts the Span Check cycle. The operation is similar to the Span 

Calibration cycle above, but BioTector does not update any Span Adjust values at the end of the span 

check cycle. Span Check reactions have the prefix “SK”.  

 

13. Span Program. 

Span Program

 

function allows the user to program the number of span reactions to be 

carried out during the Span Calibration and Span Check cycles. It is recommended not to modify the 

factory set Span Program value unless it is absolutely necessary. Any unnecessary modification in this 

setting may have an impact on the span adjust values.  

 

14. Span Average. 

Span Average function allows the user to program the number of reactions to be 

averaged at the end of the Span Calibration and Span Check cycles. It is recommended not to modify the 

factory set Span Program value unless it is absolutely necessary. Any unnecessary modification in this 

setting may have an impact on the span adjust values.  

 

15.  Range. 

Range function allows the user to program the operation range at which the Span Calibration and 

Span Check reactions are carried out. If the selected range is in conflict with the programmed 

CALIBRATION STANDARD concentration in this menu, system automatically displays a “Caution! 

Reaction range or Standard is Incorrect” warning. See System Range Data screen (see Sectio

2.2.3

 

System Range Data Screen

  for details)  in order to select the correct operation range or correct 

standard solution.   

 

16. TOC Cal Std.

 TOC Calibration Standard function allows the user to program the concentration (mgC/l) of 

the TOC standard solution used in Span Calibration reactions. If the programmed concentration level is in 

conflict with the programmed RANGE above, system automatically displays a “Caution! Reaction Range 

or Standard is Incorrect” warning. See System Range Data screen in order to select the correct operation 

range or correct standard solution. If TOC Calibration Standard is programmed as 0.0mgC/l, system does 

not calculate or update any Span Adjust factors and therefore omits any span related warnings defined 

above. See Section 

6.2 

Calibration Standards

  for the details of BioTector standard solutions and 

preparation procedures. In TC systems, this parameter is named as TC Calibration Standard. 

 

17. TOC Check Std.

 TOC Check Standard function allows the user to program the concentration (mgC/l) of 

the TOC standard solution used in Span Check reactions. If TOC Check Standard  is programmed as 

0.0mgC/l, any span related warnings are omitted. In TC systems, this parameter is named as TC Check 

Standard. 

 

18. TIC Cal Std.

 TIC Calibration Standard function allows the user to program the concentration (mgC/l) of 

the TIC standard solution used in Span Calibration reactions. If TIC Calibration Standard is programmed 

as 0.0mgC/l, any span related warnings are omitted. In VOC and TC – TIC systems, it is recommended to 

run the TIC and TOC calibrations separately using separate standard solutions. 

 

19. TIC Check Std.

 TIC Check Standard function allows the user to program the concentration (mgC/l) of the 

TIC standard solution used in Span Check reactions. If TIC Check Standard is programmed as 0.0mgC/l, 

any span related warnings are omitted. 

 

Summary of Contents for BioTector B3500

Page 1: ...DOC023 53 90740 BioTector B3500 Online TOC Analyzer BASIC USER MANUAL 6 2018 Edition 1...

Page 2: ......

Page 3: ...cautions 12 1 5 3 Pr cautions chimiques 12 1 5 4 Sample Stream Precautions 13 1 5 4 Pr cautions relatives aux chantillons 14 SECTION 2 OPERATOR S MANUAL 15 2 1 SOFTWARE SCREENS AND SOFTWARE MENU DIAGR...

Page 4: ...be impaired 1 1 Information and Safety Signs used in the Manual When any supplementary information is required and if any hazards exist the necessary information and safety signs Information Caution...

Page 5: ...sary to work with this component it should be handled with care This symbol when noted on a product illustrates the risk of chemical harm due to its corrosive acidic caustic or solvent nature Only qua...

Page 6: ...rop ene indicates that The instrument complies with the European product directives health safety and environmental protection legislations Conforms to ANSI UL Std 61010 1 Certified to CAN CSA Std 610...

Page 7: ...ned herein or for incidental or consequential damages in connection with the furnishing performance or use of this material The information contained in this manual is subject to change without notice...

Page 8: ...ary oedema may develop with delayed onset usually one or more hours after exposure Following severe acute ozone toxification recovery is slow In the few severe human cases reported 10 14 days of hospi...

Page 9: ...s g n rales de s curit Pri re d tre toujours attentif toutes les notices de prudence d avertissement ou de danger Le non respect des instructions de s curit peut engendrer la blessure grave d individu...

Page 10: ...en Isoler les fils lectriques de l appareil avant de commencer tout travail dans le bo tier lectronique Seul le personnel lectricien qualifi est habilit effectuer tous travaux d lectricit Se conformer...

Page 11: ...t Pour son fonctionnement BioTector emploie de l oxyg ne O2 comme gaz porteur L oxyg ne ne doit comporter aucun gaz carbonique CO2 ni d azote N2 Le taux moyen de consommation d oxyg ne du BioTector es...

Page 12: ...sing appropriate safety equipment 1 5 3 Pr cautions chimiques La liste de la Section 6 R actifs et Standards de Calibration Section 6 Reagents and Calibration Standards num re un certain nombre de pro...

Page 13: ...FA Per fluoro alkoxy Vinyl Reinforced Silicon Reinforced PVC Poly vinyl chloride Fittings PFA Per fluoro alkoxy Stainless Steel SS 316 PVDF Poly vinylidene flouride Brass Pump Tubing EMPP Elastomer mo...

Page 14: ...xy Vinyle Silicone renforc PVC renforc poly chlorure de vinyle Raccords PFA perfluoroalkoxy Acier inoxydable SS 316 PVDF polyfluorure de polyvinylid ne Laiton Tuyauterie de la pompe PPMOD polypropyl n...

Page 15: ...ramming the BioTector The ENTER key which advances user to the next screen is also used to enter programmed settings in the BioTector The symbols used on BioTector LCD screen and their meanings are as...

Page 16: ...Page 16 Software Menu Diagram...

Page 17: ...yed on the top left hand side of the Analysis Data and Reagent Status screens On most other screens only the screen name is displayed in this location System status messages are displayed in the follo...

Page 18: ...mple if the reaction is currently in the TIC Base Oxidation TOC phase The operation Range e g Range 1 2 or 3 the BioTector is using to carry out its analysis The Reaction Time which is the elapsed tim...

Page 19: ...he analyzer in C e g 26 C The CO2 instantaneous value e g 56ppm CO2i and the CO2 zero value e g 12ppm CO2z of the reaction The elapsed time e g 265s since the start of the analysis 2 1 5 Reagent Statu...

Page 20: ...r diagnostics to download data to program the software functions and to program the system specific settings in the BioTector The sub menus in this level can be password protected using the Password m...

Page 21: ...Standby in the system reaction archive The 4 20mA signal or other output devices are not updated When remote standby signal is deactivated the BioTector starts analyzing When remote standby signal is...

Page 22: ...eyboard the BioTector stops as soon as its present reaction is completed An is displayed to let the operator know the function has been activated 4 Emergency Stop When this function is activated the B...

Page 23: ...ils The comprehensive Zero Check Calibration and Span Check Calibration parameters are programmed in Zero Program and Span Program menus respectively see Section 8 3 4 2 Zero Program and 8 3 4 3 Span...

Page 24: ...n Check menu 3 If Span Calibration or Span Check is activated in New Reagents Program menu a Span Calibration or Span Check is carried out Once the procedure is completed the BioTector either stops or...

Page 25: ...cate that this function has been selected If the BioTector is stopped then the Manual mode will start immediately To deactivate this function before the manual operation mode has started press the ENT...

Page 26: ...on Archive screens To access the screens press the ENTER key and to return to the previous screen press the ESCAPE key 2 2 6 Fault Archive Menu In the Fault Archive menu the user can view the last 99...

Page 27: ...startup time for a scheduled task for example the startup time for a manual sample sequence in Manual Program menu 2 2 8 Contact Information Contact Information menu displays the manufacturer distribu...

Page 28: ...em only checks the zero response at each range and displays the suggested Zero Adjust values in brackets for all ranges next to the current Zero Adjust settings When a Zero Check cycle is completed if...

Page 29: ...8 Zero Program Zero Program function allows the user to program the number of zero reactions run at one or more ranges R1 R2 and or R3 When the number of zero calibration reactions for one or two of t...

Page 30: ...below are displayed in this menu 1 3 TOC Span Adjust This menu item allows the user to set the TOC span adjust factors manually by entering the STANDARD solution used and the calibrated average react...

Page 31: ...alibration and Span Check reactions are carried out If the selected range is in conflict with the programmed CALIBRATION STANDARD concentration in this menu system automatically displays a Caution Rea...

Page 32: ...for TOC Calibration Standard above Section 8 3 4 3 Span Program and Section 6 2 Calibration Standards for further details 21 TC Check Std In VOC systems the sum of the TIC and TOC Check Standard solut...

Page 33: ...DOC012 97 80506 User Guidance for EMC Class A Equipment EMC A A A...

Page 34: ...years The product manufacturing date can be found on the product label on the main board enclosure behind the display BioTector 25 EFUP CHINA ROHS DECLARATION TABLE ROHS Part name Toxic or Hazardous...

Page 35: ......

Page 36: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de...