Amperometric Titration Theory
50081_Titration.fm
Page 145
Appendix A Amperometric Titration Theory
1.1 General Titration Theory
All titrations are carefully regulated and monitored chemical reactions. Any chemical
reaction involves adding one reactant to another to form one or more reaction products.
1.1.1 Redox Reactions
Oxidation-reduction reactions (Redox reactions) are a special class of chemical reactions
where electrons are transferred from one species to another to form new chemical
species. Species gaining electrons are said to be reduced while species releasing
electrons are said to be oxidized. Species accepting electrons are called oxidizing agents
(or oxidants) and those releasing electrons are called reducing agents (or reductants).
Chlorine (Cl
2
) is a common oxidizing reagent because it can break into separate chlorine
atoms (Cl) that readily accept electrons from other species. Another common oxidizing
agent is iodine (I
2
). phenylarsine oxide (abbreviated PAO representing PhAsO, where
Ph = C
6
H
5
–
) and sodium thiosulfate (Na
2
S
2
O
3
) are examples of common reductants.
1.1.2 Reaction Measurement
The progress of an oxidation-reduction reaction can be monitored by measuring the
electrical current passing through a solution under an imposed potential when electrons
are transferred from the reductant to the oxidant. Current stops flowing if either the
oxidant or reductant is consumed or is otherwise removed from the solution.
In an amperometric titration a titrant solution containing a reductant is added to a sample
containing an oxidant (or a oxidant is added to a sample containing an reductant) while
the electrical current passing between two platinum electrodes is monitored. The species
being analyzed is called the analyte. The point at which equal amounts of oxidant and
reductant are brought together is called the end point. The end point in an amperometric
titration curve is identified as that point in the titration curve (electrical current vs. titrant
volume) where a sharp break (i.e., change in slope) occurs in the titration curve.
Depending on the oxidation-reduction reaction, the titration can be either a forward or
back titration.
1.1.2.1 Forward vs. Back
In a forward titration (
), the measured current starts at a high initial value. It
drops continuously as titrant is added until the end point is reached and the current levels
off. In a back titration (
) the measured current starts at a low initial value and
remains level as titrant is added. When the end point is reached the current increases
continuously as titrant is added.
Reactant 1
(
)
Reactant 2
(
)
Reaction Products
(
)
→
+
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......