Amperometric Titration Theory
Amperometric Titration Theory
50081_Titration.fm
Page 147
1.1.3.2 Sample Spike
In a back titration (
), the sample solution is spiked with a measured volume of
standard reductant solution (e.g., phenylarsine oxide or sodium thiosulfate). This fixes
the chlorine concentration, allowing the sample to be stored for later analysis. The
chlorine concentration is derived from the difference between the amounts of reductant
originally added to the sample (N
Reductant Spike
V
Reductant Spike
) and that remaining after
reacting with the Chlorine in the sample (N
Titrant
V
Titrant
). The amount of unreacted
reductant is determined by titration with a standard iodine solution.
Figure 2
Total Chlorine Back Titration
1.2 Errors and Interferences
1.2.1 Overview
According to
Standard Methods,
“the amperometric method is the method of choice
because it is not subject to interference from color, turbidity, iron, manganese, or nitrite-
nitrogen”. However, several of these factors may affect the determination of chlorine
species when using amperometric methods. This is a brief review of common sources of
error encountered with actual samples.
The common chlorine methods will detect other disinfectants such as chlorine dioxide
(ClO
2
), ozone (O
3
), bromine (Br
2
), hydrogen peroxide (H
2
O
2
) and disinfectant by-
products such as chlorite and chlorate. In the free chlorine determinations, these oxidants
will be reduced to varying degrees by PAO or thiosulfate in the titration method. Each of
these oxidants will oxidize iodide to iodine, interfering in the total chlorine
determination.
1.2.2 Deposition on Electrode Surfaces
The AutoCAT 9000 is designed with self-cleaning platinum electrodes. Occasionally, it
may be necessary to perform the embedded clean/condition procedure to maintain
optimal performance. Sharp amperometric titration end points require clean, well-
C
sample
N
Reductant Spike
V
Reductant Spike
N
Titrant
V
Titrant
–
V
Sample
---------------------------------------------------------------------------------------------------------------------------
35453 mg/equ wt.
×
=
Titrant Volume (mL)
Current (µAmp)
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......