![Hach AutoCAT 9000 Instruction Manual Download Page 148](http://html1.mh-extra.com/html/hach/autocat-9000/autocat-9000_instruction-manual_3154724148.webp)
Amperometric Titration Theory
Amperometric Titration Theory
Page 146
50081_Titration.fm
Figure 1
Total Chlorine Forward Titration
1.1.3 Determining Concentration
Experimentally the concentration of the titrant is known very accurately, as is the volume
of titrant that must be added to reach the titration end point. Use the following formula to
calculate the concentration of the analyte in an unknown sample:
Where:
V
Sample
is the sample volume
V
Titrant
is the volume of titrant solution that must be added to reach the end point
N
Titrant
is the concentration (normality) of the titrant solution
N
Sample
is the concentration (normality) of the analyte in the sample solution
1.1.3.1 Conversion Factor
Use a conversion factor to convert normality to more tangible units such as mg/L or ppm.
The precise conversion factor depends on the particular analyte and concentration units
involved. For example, one equivalent weight of chlorine (Cl
2
) is 35.453 g.
Consequently, a 0.100 N chlorine solution contains 3.5453 grams of Chlorine per liter of
solution. The following conversion factors can be used to convert from normality to the
more common concentration units.
Inserting the conversion factor into the equation permits the chlorine concentration to be
expressed in (mg/L), as shown below:
The above equation applies to
forward
titrations only. A slightly different calculation,
shown below, is used for
back
titrations.
Titrant Volume (mL)
Current (µAmp)
NSample
VTitrantNTitrant
VSample
-------------------------------------------
=
Chlorine Concentration: (mg/L)
(ppm)
N
chlorine
35453 mg/eq wt.
×
=
=
C
sample
N
Reductant Spike
V
Reductant Spike
V
Sample
-----------------------------------------------------------------------------
35453 mg/equ wt.
×
=
Summary of Contents for AutoCAT 9000
Page 2: ...50081_TTL fm...
Page 8: ...Page 6 50081_Safety fm...
Page 14: ...Page 12 50081_Introduction fm...
Page 24: ...Page 22 50081_OperationStop fm...
Page 34: ...Basic Operation Basic Operation Page 32 50081_Operation fm...
Page 44: ...Page 42 50081_SoftwareSetup fm...
Page 51: ...Calibration Procedures 50081_CalibrationStopper fm Page 49 Section 6 Calibration Procedures...
Page 52: ...Page 50 50081_CalibrationStopper fm...
Page 66: ...Page 64 50081_Sampling fm...
Page 67: ...Titration Procedures 50081_ProcedureStopper fm Page 65 Section 8 Titration Procedures...
Page 68: ...Page 66 50081_ProcedureStopper fm...
Page 138: ...Page 136 50081_ManualEndpoint fm...
Page 140: ...Page 138 50081_BDL fm...
Page 154: ...Page 152 50081_Titration fm...
Page 158: ...Page 156 50081_PrinterInstallation fm...
Page 160: ...Page 158 50081_ReplacementParts fm...
Page 170: ...Page 168 50081_GeneralInfoStop fm...
Page 176: ......