Page 38
Additional Network Menu Features
51350-18 Initial Aquatrend Setup.fm
Section 3
3.9.1.3 Displaying Recorder to Analog Out Information
The information identifies which sensor alarm is attached to each individual
SOM recorder.
1.
Press the MENU key to access the Main Menu.
2.
Move the pointer to Network Menu and press the ENTER key.
3.
Select REVIEW NET CONNECTIONS and press the ENTER key.
4.
Use the up or down arrow keys to move the pointer to
RECORDER TO ANALOG OUT and press the ENTER key. All recorder
to analog out connections will be displayed.
3.9.1.4 Sensor to Display
The sensor to display connections identify which sensor is attached
(via the software) to which Digital Display Module (DDM). The setup
procedure for the DDM allows the user to send the data from any sensor to
any DDM on the system, even if the DDM is not physically connected to that
sensor. To determine which sensor the DDM is attached to (via the software),
proceed as follows:
1.
Press the MENU key to access the Main Menu.
2.
Move the pointer to NETWORK MENU and press the ENTER key.
3.
Select REVIEW NET CONNECTIONS and press the ENTER key.
4.
Select SENSOR TO DDM and press ENTER.
Sensor to DDM connection information will be displayed.