English
7
Operation
Calibration (continued)
Sensor code
Entry of the sensor code
Lab value correction
The laboratory values of the last matrix correction can be changed
Enter the lab value.
Enter the laboratory values if 1-point matrix correction or 2-point matrix correction has
been selected
Ammonium
1,2
Entry of ammonium laboratory value
Nitrate
1,3
Enter the laboratory value for nitrate
Potassium
1,2
Enter the laboratory value for potassium
Chloride
1,3
Enter the laboratory value for chloride
Confirm the entry.
Confirm the values entered
Correction result
Display the correction results
NH
4
–N
1,2
Displays whether or not the ammonium correction was successful
NO
3
–N
1,3
Shows whether or not the nitrate correction was successful
K+
1,2
Displays whether or not the potassium correction was successful
Cl
1,3
Displays whether or not the chloride correction was successful
Information
Information on the matrix correction used per parameter
NH
4
–N
1,2
Matrix correction used for ammonium
NO
3
–N
1,3
Matrix correction used for nitrate
K+
1,2
Matrix correction used for potassium
Cl
1,3
Matrix correction used for chloride
Settings
Name
Enter or edit the name.
Up to 10 alphanumeric characters
Unit
Select either mg/L or ppm as the measurement unit
Parameter
Select NH
4
–N or NH
4
and/or NO
3
–N or NO
3
Temperature
Selection of °C or °F as temperature unit
Offset temperature
Enter a temperature offset
Response time
Entry of the response time (30 sec to 300 sec)
Datalog interval
Select the data log interval (OFF, 30 sec, 1 min, 2 min, 5 min, 10 min, 15 min and
30 min), 5 min is the factory setting
K+ compensation
1,2
Select automatic potassium compensation:
On
Off
0 = compensation OFF
0.1–2000 mg/L Cl = Fixed compensation value
K+ concentration
1,2
Only shown when K+ COMPENSATE is
OFF
Cl - compensation
1,3
Select automatic chloride compensation:
On
Off
0 = compensation OFF
0.1–2000 mg/L Cl = Fixed compensation value
Cl - concentration
1,3
Only shown when Cl COMPNSATE is
OFF
Reset configuration
Resets the configuration to the factory setting
Maintenance
Sensor information
Information on the connected sensor
Sensor type
Name of the connected sensor
4.2 Sensor setup menu (Continued)
Summary of Contents for AISE sc
Page 2: ......
Page 4: ...4 Table of contents...
Page 78: ...78 Fran ais D pannage...