Table 1  Display icons

Icon Description

Shows the battery power level. Flashes when the battery power is less than 10%.

Shows when the lock feature is on. When the lock feature is on, the parameter value on the display does not change.
Push   to set the lock feature to on or off.

Shows when the tester is in calibration mode.

Keypad description

Figure 5

 shows the keypad. 

Table 2

 gives the key descriptions.

Figure 5  Keypad overview

1   

Power key

2   

Calibration/Settings key

3   

Lock key

Table 2  Key functions

Key

Description

Push and hold to set the power to on or off.

Push to start a calibration. To exit a calibration, push and hold.
Push and hold until "SEt" shows on the display to go to the settings menu. To exit the settings menu, push and hold
until "End" shows on the display. When in the settings menu, push to scroll through the settings.

Note: Power cannot be set to off while in settings or calibration mode.

Push to set the lock feature to on or off. When the lock feature is on, the lock icon shows and the parameter value on
the display does not change.

Calibration

C A U T I O N

 

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment
appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.

Calibrate the tester before initial use and when:

Results drift

Results are not accurate

Before the initial calibration and after a dry storage, soak the sensor for several minutes in the sample or tap
water.

6

Summary of Contents for 9531600

Page 1: ...User Manual DOC022 53 80400 Pocket Pro Salinity 1...

Page 2: ...s it describes at any time without notice or obligation Revised editions are found on the manufacturer s website Safety information The manufacturer is not responsible for any damages due to misapplic...

Page 3: ...e complies with Part 15 of the FCC Rules Operation is subject to the following conditions 1 The equipment may not cause harmful interference 2 The equipment must accept any interference received inclu...

Page 4: ...ged contact the manufacturer or a sales representative immediately Figure 2 Product components 1 Pocket Pro tester 3 Lanyard 2 AAA alkaline batteries 4x Install the batteries CAU TI ON Explosion hazar...

Page 5: ...erface and navigation Display description Figure 4 shows the measurements calibration standard information and indicator icons shown on the display Table 1 gives descriptions of the icons Figure 4 Dis...

Page 6: ...on the display to go to the settings menu To exit the settings menu push and hold until End shows on the display When in the settings menu push to scroll through the settings Note Power cannot be set...

Page 7: ...ater sample into the cap to the fill line 6 Put the sensor fully into the cap The measured value shows on the top line 7 To keep the measured value on the display when the sensor is removed from the s...

Page 8: ...e measurement range of the tester Gently clean the sensor Refer to Clean the sensor on page 8 Calibrate as necessary bottom line The temperature value is not within the operating temperature range of...

Page 9: ...Table 4 Replacement parts Description Quantity Item no AAA alkaline batteries 4 pkg 4674300 Lanyard 1 201305 SINGLET 12 88 mS cm 20 pkg 2771520 9...

Page 10: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1222 V sen...

Reviews: