W A R N I N G

 

Fire hazard. Battery substitution is not permitted. Use only alkaline batteries.

Install the four AAA alkaline batteries in the tester. Refer to the illustrated steps in 

Figure 3

.

Figure 3  Install the batteries

User interface and navigation

Display description

Figure 4

 shows the measurements, calibration standard information and indicator icons shown on the display. 

Table 1

 gives descriptions of the icons.

Figure 4  Display overview

1   

Calibration icon

4   

Battery icon

7   

Calibration standard expected

1

2   

Lock icon

5   

Parameter value

8   

Temperature

3   

Parameter

6   

Calibration standard measured for
last calibration (high)

1

Shows during calibration

5

Summary of Contents for 9531600

Page 1: ...User Manual DOC022 53 80400 Pocket Pro Salinity 1...

Page 2: ...s it describes at any time without notice or obligation Revised editions are found on the manufacturer s website Safety information The manufacturer is not responsible for any damages due to misapplic...

Page 3: ...e complies with Part 15 of the FCC Rules Operation is subject to the following conditions 1 The equipment may not cause harmful interference 2 The equipment must accept any interference received inclu...

Page 4: ...ged contact the manufacturer or a sales representative immediately Figure 2 Product components 1 Pocket Pro tester 3 Lanyard 2 AAA alkaline batteries 4x Install the batteries CAU TI ON Explosion hazar...

Page 5: ...erface and navigation Display description Figure 4 shows the measurements calibration standard information and indicator icons shown on the display Table 1 gives descriptions of the icons Figure 4 Dis...

Page 6: ...on the display to go to the settings menu To exit the settings menu push and hold until End shows on the display When in the settings menu push to scroll through the settings Note Power cannot be set...

Page 7: ...ater sample into the cap to the fill line 6 Put the sensor fully into the cap The measured value shows on the top line 7 To keep the measured value on the display when the sensor is removed from the s...

Page 8: ...e measurement range of the tester Gently clean the sensor Refer to Clean the sensor on page 8 Calibrate as necessary bottom line The temperature value is not within the operating temperature range of...

Page 9: ...Table 4 Replacement parts Description Quantity Item no AAA alkaline batteries 4 pkg 4674300 Lanyard 1 201305 SINGLET 12 88 mS cm 20 pkg 2771520 9...

Page 10: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1222 V sen...