44
sc1000 Operation
7.
Press
ENTER
when stable. A screen will display 1 Point Auto Complete and the slope
(XX.X mV/pH).
8.
Return the probe to process.
5.7.4 Two Point Manual Calibration Method
1.
Tap the bottom-left of the screen to display the pop-up toolbar and press
MAIN MENU
.
2.
From the Main Menu, select SENSOR SETUP and press
ENTER
.
3.
Select the appropriate sensor if more than one is attached and press
ENTER
.
4.
Select CALIBRATE and press
ENTER
.
5.
Select 2 POINT MANUAL CAL and tap the dark-blue highlighted area to the far-right.
Select the available Output Mode (Active, Hold, or Transfer) from the list box and
press
ENTER
.
6.
Move the clean probe to Solution 1 and press
ENTER
to confirm.
7.
Press enter when stable. Edit the solution value using the keypad and press
ENTER
.
8.
Move probe to solution 1 and press
ENTER
.
9.
Press enter when stable. Edit the solution value using the keypad and press
ENTER
.
10. A screen will display 2 Point Manual Cal Complete and the slope (XX.X mV/pH).
11. Return the probe to process.
5.7.5 One Point Manual Calibration Method
1.
Tap the bottom-left of the screen to display the pop-up toolbar and press
MAIN MENU
.
2.
From the Main Menu, select SENSOR SETUP and press
ENTER
.
3.
Select the appropriate sensor if more than one is attached and press
ENTER
.
4.
Select CALIBRATE and press
ENTER
.
5.
Select 1 POINT MANUAL and tap the dark-blue highlighted area to the far-right.
Select the available Output Mode (Active, Hold, or Transfer) from the list box and
press
ENTER
.
6.
Move the clean probe to solution and press
ENTER
to continue.
7.
Press enter when stable. Edit the solution value using the keypad and press
ENTER
.
8.
Press
ENTER
when stable. A screen will display 1 Point Manual Complete and the
slope (XX.X mV/pH).
9.
Return the probe to process.
Summary of Contents for 8362sc
Page 2: ...Visit http www hach com...
Page 6: ...Visit us at www hach com...
Page 10: ...Visit us at www hach com...
Page 14: ...14 Installation Figure 3 Measurement System Electrical Connection to Digital Gateway...
Page 16: ...Visit us at www hach com...
Page 31: ...31 sc100 Operation Figure 8 Overflow Sampling...
Page 47: ...47 sc1000 Operation Figure 16 Overflow Sampling...
Page 50: ...Visit us at www hach com...
Page 52: ...Visit us at www hach com...
Page 54: ...Visit us at www hach com...
Page 58: ...Visit us at www hach com...