9
Section 3
Installation
DANGER
Only qualified personnel should conduct the installation tasks described in this
section of the manual.
3.1 Connecting/Wiring the Sensor to the sc100
The sensor cable is supplied with a keyed quick-connect fitting for easy attachment to the
controller, see
. Retain the connector cap to seal the connector opening in case
the sensor must be removed. Optional extension cables may be purchased to extend the
sensor cable length. If the total cable length exceeds 100 m (300 ft), a termination box
must be installed. See
.
Figure 1
Attaching the Sensor using the Quick-connect Fittings
3.1.1 Hard-wiring the Sensor to the sc100 Controller
1.
Open the controller cover.
2.
Disconnect and remove the existing wires between the quick connect and terminal
strip J5, see
.
3.
Remove the quick-connect fitting and wires and install the threaded plug on the
opening to maintain the environmental rating.
4.
Cut the connector from the sensor cable.
5.
Strip the insulation on the cable back 1-inch. Strip ¼-inch of each individual wire end.
6.
Pass the cable through conduit and a conduit hub or a strain relief fitting and an
available access hole in the controller enclosure. Tighten the fitting.
7.
Reinstall the plug on the sensor access opening to maintain the environmental rating.
8.
Wire as shown in
.
9.
Close and secure the controller cover.
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY A
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+ DATA
+ DATA
+ OUT 2
+ OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+ V
+ V
+ OUT 1
+ OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY A
J1
J2
J4
S1
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
1
1
+ DATA
+ DATA
+ OUT 2
+ OUT 2
– DATA
– OUT 2
SERVICE REQUEST
SHIELD/CHASSIS GND
+ V
+ V
+ OUT 1
+ OUT 1
GND
– OUT 1
2
2
3
3
4
4
5
5
6
PROBES
ANAL
OG OUTP
UTS
ANAL
OG OUTP
UTS
PCB
CONNECTOR
PCB
CONNECTOR
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
FIELD WIRING
INSULATION MUST
BE RATED TO
80° C MINIMUM
DANGER - EXPLOSION HAZARD
DANGER - RISQUE D'EXPLOSION
DO NOT DISCONNECT WHILE CIRCUIT IS LIVE
UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.
NE PAS DEBRANCHER TANT QUE LE EST SOUS
TENSION, A MONIS QU'IL NE S'AGISSE D'UN
EMPLACEMENT NON-DANGEROUX
J5
J5
客服:400 688 9809
Summary of Contents for 5740sc
Page 1: ...Visit http www hach com 客服 400 688 9809 ...
Page 6: ...Visit us at www hach com 客服 400 688 9809 ...
Page 12: ...12 Installation Figure 4 Sensor Installation Examples 客服 400 688 9809 ...
Page 22: ...Visit us at www hach com 客服 400 688 9809 ...
Page 34: ...Visit us at www hach com 客服 400 688 9809 ...
Page 38: ...Visit us at www hach com 客服 400 688 9809 ...
Page 40: ...Visit us at www hach com 客服 400 688 9809 ...
Page 44: ...Visit us at www hach com 客服 400 688 9809 ...